Home › HITACHI HIGH-TECH CORPORATION › PROBEAT
PROBEAT
HITACHI HIGH-TECH CORPORATION · Model PROBEAT WITH DISCRETE SPOT SCANNING SYSTEM · UDI-DI 04560333350358
LHN
Prescription
1
510(k) clearances
136
Adverse events (2 deaths)
58
Recalls
27
Facilities
Description
The PROBEAT with DSSS is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. The only difference between the PROBEAT with DSSS and the cleared PROBEAT is the addition of the Discrete Spot Scanning System ("DSSS").
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2026-02-06 | Hitachi, Ltd. Radiation Oncology Systems, Kashiwanoha | Software anomaly in the patient positioning system may result in positional discrepancy. |
| 2026-02-06 | Hitachi, Ltd. Radiation Oncology Systems, Kashiwanoha | Software anomaly in the patient positioning system may result in positional discrepancy. |
| 2026-02-03 | Ion Beam Applications S.A. | It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the |
| 2024-07-10 | Ion Beam Applications S.A. | Temporary deactivation of collision detection for Forte Robotic Patient Positioning System occasionally experiences issues |
| 2024-07-08 | Ion Beam Applications S.A. | Safety Parameters Verification Mechanisms can be deactivated in clinical runtime |
| 2024-05-27 | Hitachi America, Ltd., Particle Therapy Division | During positioning within the system's 3D to 3D matching mode while transferring CBCT images, when the Float Image Selection button is pressed, it may display CBCT images that are |
| 2024-05-27 | Hitachi America, Ltd., Particle Therapy Division | During positioning within the system's 3D to 3D matching mode while transferring CBCT images, when the Float Image Selection button is pressed, it may display CBCT images that are |
| 2024-05-27 | Hitachi America, Ltd., Particle Therapy Division | During positioning within the system's 3D to 3D matching mode while transferring CBCT images, when the Float Image Selection button is pressed, it may display CBCT images that are |
| 2024-05-10 | Ion Beam Applications S.A. | Irradiation is not interrupted when some C230 Accelerator Control Unit critical parameters are outside of the defined tolerances |
| 2024-03-04 | Ion Beam Applications S.A. | There is a risk of mistreatment as irradiation is not prevented when some safety parameters are out of tolerance. |