FDA Device Intelligence
HomeDevice types › LHN

System, Radiation Therapy, Charged-Particle, Medical

FDA product code LHN · Class 2 · Radiology

226
Registered devices
136
Adverse events
58
Recalls
92
510(k) clearances

Devices with this code

Recent recalls

DateFirmReason
2026-02-06Hitachi, Ltd. Radiation Oncology Systems, KashiwanohaSoftware anomaly in the patient positioning system may result in positional discrepancy.
2026-02-06Hitachi, Ltd. Radiation Oncology Systems, KashiwanohaSoftware anomaly in the patient positioning system may result in positional discrepancy.
2026-02-03Ion Beam Applications S.A.It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS do
2024-07-10Ion Beam Applications S.A.Temporary deactivation of collision detection for Forte Robotic Patient Positioning System occasionally experiences issues
2024-07-08Ion Beam Applications S.A.Safety Parameters Verification Mechanisms can be deactivated in clinical runtime
2024-05-27Hitachi America, Ltd., Particle Therapy DivisionDuring positioning within the system's 3D to 3D matching mode while transferring CBCT images, when the Float Image Selection button is pressed, it may display C
2024-05-27Hitachi America, Ltd., Particle Therapy DivisionDuring positioning within the system's 3D to 3D matching mode while transferring CBCT images, when the Float Image Selection button is pressed, it may display C
2024-05-27Hitachi America, Ltd., Particle Therapy DivisionDuring positioning within the system's 3D to 3D matching mode while transferring CBCT images, when the Float Image Selection button is pressed, it may display C
2024-05-10Ion Beam Applications S.A.Irradiation is not interrupted when some C230 Accelerator Control Unit critical parameters are outside of the defined tolerances
2024-03-04Ion Beam Applications S.A.There is a risk of mistreatment as irradiation is not prevented when some safety parameters are out of tolerance.

Query this across every FDA database — by API.

Linked GUDID · 510(k)/PMA · MAUDE · recalls · establishments. Free tier, no card required.

Get your free API keyRead the docs
Sourced from public U.S. FDA data. Informational only — not medical or regulatory advice. Not affiliated with the FDA.