PROBEAT
HITACHI, LTD. · Model Small Field Applicator for PROBEAT-V · UDI-DI 04560333350280
Description
The PROBEAT-V is a proton beam irradiation system, which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam with the prescribed dose, dose distribution and directed to the prescribed patient treatment site. Small Field Applicator is an optional accessory to the proton beam therapy system which can be added to the nozzle configuration of the cleared PROBEAT-V system to make the lateral penumbra sharp, as needed. The Small Field Applicator may be used in place of the optional removable Applicator having an aperture (collimator) that has been cleared as part of the PROBEAT-V system.
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2026-02-06 | Hitachi, Ltd. Radiation Oncology Systems, Kashiwanoha | Software anomaly in the patient positioning system may result in positional discrepancy. |
| 2026-02-06 | Hitachi, Ltd. Radiation Oncology Systems, Kashiwanoha | Software anomaly in the patient positioning system may result in positional discrepancy. |
| 2026-02-03 | Ion Beam Applications S.A. | It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the |
| 2024-07-10 | Ion Beam Applications S.A. | Temporary deactivation of collision detection for Forte Robotic Patient Positioning System occasionally experiences issues |
| 2024-07-08 | Ion Beam Applications S.A. | Safety Parameters Verification Mechanisms can be deactivated in clinical runtime |
| 2024-05-27 | Hitachi America, Ltd., Particle Therapy Division | During positioning within the system's 3D to 3D matching mode while transferring CBCT images, when the Float Image Selection button is pressed, it may display CBCT images that are |
| 2024-05-27 | Hitachi America, Ltd., Particle Therapy Division | During positioning within the system's 3D to 3D matching mode while transferring CBCT images, when the Float Image Selection button is pressed, it may display CBCT images that are |
| 2024-05-27 | Hitachi America, Ltd., Particle Therapy Division | During positioning within the system's 3D to 3D matching mode while transferring CBCT images, when the Float Image Selection button is pressed, it may display CBCT images that are |
| 2024-05-10 | Ion Beam Applications S.A. | Irradiation is not interrupted when some C230 Accelerator Control Unit critical parameters are outside of the defined tolerances |
| 2024-03-04 | Ion Beam Applications S.A. | There is a risk of mistreatment as irradiation is not prevented when some safety parameters are out of tolerance. |