Recalls

DateFirmReason
2024-04-09QUIDELORTHOT3 Reagent Packs used with Free T3 Calibrators may produce positively biased results for patient and quality control samples. The positive bias may cause erroneous patient results
2023-03-15Beckman Coulter, Inc.The reagent lot produces elevated results (approximately >30%) in some (15-20%)patient samples that do not correlate with the results of other thyroid markers nor Free T3 results g
2019-04-30Beckman Coulter Inc.The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays: Free T3-REF A13422
2019-04-30Beckman Coulter Inc.The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays: Free T3-REF A13422
2018-11-30Tosoh Bioscience Inc
2018-10-12Abbott Ireland Diagnostics DivisionSamples tested using ARCHITECT Free T3 or ARCHITECT Total T3 assays may show depressed results due to a reagent carryover when testing on board with specific assays on the ARCHITEC
2018-10-12Abbott Ireland Diagnostics DivisionSamples tested using ARCHITECT Free T3 or ARCHITECT Total T3 assays may show depressed results due to a reagent carryover when testing on board with specific assays on the ARCHITEC
2018-03-05Tosoh Bioscience IncAsfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, depending upon the assay.
2018-03-05Tosoh Bioscience IncAsfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, depending upon the assay.
2017-07-28Siemens Healthcare Diagnostics, IncSiemens has identified an issue when scanning the 20 barcode for the Master Curve Card (MCC) with ADVIA Centaur systems T3 kit lots ending in 198 on the ADVIA Centaur XPT system.