FDA Device Intelligence
HomeDevice types › CDP

Radioimmunoassay, Total Triiodothyronine

FDA product code CDP · Class 2 · Clinical Chemistry

123
Registered devices
1,979
Adverse events
20
Recalls
144
510(k) clearances

Devices with this code

Recent recalls

DateFirmReason
2024-04-09QUIDELORTHOT3 Reagent Packs used with Free T3 Calibrators may produce positively biased results for patient and quality control samples. The positive bias may cause errone
2023-03-15Beckman Coulter, Inc.The reagent lot produces elevated results (approximately >30%) in some (15-20%)patient samples that do not correlate with the results of other thyroid markers n
2019-04-30Beckman Coulter Inc.The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays
2019-04-30Beckman Coulter Inc.The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays
2018-11-30Tosoh Bioscience Inc
2018-10-12Abbott Ireland Diagnostics DivisionSamples tested using ARCHITECT Free T3 or ARCHITECT Total T3 assays may show depressed results due to a reagent carryover when testing on board with specific as
2018-10-12Abbott Ireland Diagnostics DivisionSamples tested using ARCHITECT Free T3 or ARCHITECT Total T3 assays may show depressed results due to a reagent carryover when testing on board with specific as
2018-03-05Tosoh Bioscience IncAsfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, dependin
2018-03-05Tosoh Bioscience IncAsfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, dependin
2017-07-28Siemens Healthcare Diagnostics, IncSiemens has identified an issue when scanning the 20 barcode for the Master Curve Card (MCC) with ADVIA Centaur systems T3 kit lots ending in 198 on the ADVIA C

Query this across every FDA database — by API.

Linked GUDID · 510(k)/PMA · MAUDE · recalls · establishments. Free tier, no card required.

Get your free API keyRead the docs
Sourced from public U.S. FDA data. Informational only — not medical or regulatory advice. Not affiliated with the FDA.