CORTEX CPET System
Cortex Biophysik GmbH · Model MetaLyzer3B · UDI-DI 74250179233555
Description
The CORTEX CPET systems are intended for use in CPET, Spirometric, and Basal-/Resting Metabolic Rate Tests for the purpose to objectify the ventilatory, cardio-circulatory and metabolic capacity. With the following indications, they support the user in finding or confirming diagnoses, determining therapy programs and monitoring therapies. In a CPET test, a patient undergoes a period of defined physical stress (normally 10-20 minutes). Hereby data is measured about ventilation, gas exchange and heart rate. On basis of these measured data, further data is calculated and displayed, in order to depict the performance of the cardio-respiratory system. In a spirometry test, the patient performs defined breathing maneuvers (VC, FVC, MVV). Thereby measurement data of the respiratory strength and volume is collected and displayed, in order to show the mechanical capacity of the respiratory system. In a Basal-/Resting Metabolic Rate Test the patient performs a rest test (normally 15-20 minutes). Thereby measurement data of the gas exchange is collected and displayed, in order to estimate the Basal-/Resting Metabolic Rate.
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2017-08-15 | NDDd Medical Technologies, Inc. | The CO/CO2 sensor cable harness may have lower electromagnetic shielding, which might have resulted in a non-compliance with the IEC 60601-1-2 standard. As a consequence of a lowe |
| 2017-08-15 | NDDd Medical Technologies, Inc. | The CO/CO2 sensor cable harness may have lower electromagnetic shielding, which might have resulted in a non-compliance with the IEC 60601-1-2 standard. As a consequence of a lowe |
| 2014-12-23 | Shape Medical Systems, Inc | This recall has been initiated because use of the impacted DPIs could result in erroneous testing results potentially leading to incorrect diagnosis and incorrect treatment. The ma |
| 2014-05-28 | Shape Medical Systems, Inc | Shape Medical has initiated a correction due to a mandatory software upgrade for the Shape HF Cardiopulmonary Testing System prior to use of the impacted DPI lot numbers. Use of t |