Disposable Reflective Marker Spheres (DRMS) attached to reference frames and surgical instruments for use during surgical navigation in image guided surgery. One retail carton is made up of 3 Spheres per tray, 30 trays, 90 spheres total.
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2026-08-28 | Intellijoint Surgical, Inc. | A navigation system used to aid in selection and positioning of orthopedic implant system components has a workstation that may contain a 15-inch MacBook Pro battery, subject to an |
| 2026-08-20 | Think Surgical, Inc. | Robotic system square drive failed to meet design specifications potentially resulting in pin becoming stuck in square driver or the pin wobbling during drilling. |
| 2026-07-23 | Medos International Sarl | Software issue within robotic controller which may cause unintended movement of the robotic arm during bone preparation or instrument positioning. |
| 2026-05-08 | Think Surgical, Inc. | Due to manufacturing/design issue where the knob on the tracking array may loosen causing potential misalignment of the array. |
| 2026-04-23 | Kico Knee Innovation Company | The Impactor V-Block Assembly and Trackers A, B, C and E may exhibit signs of early wear and/or corrosion on the exposed magnet face. |
| 2026-04-23 | Kico Knee Innovation Company | The V-Block Assembly and Trackers A, B, C and E may exhibit signs of early wear and/or corrosion on the exposed magnet face. |
| 2026-02-10 | Orthalign, Inc | Drill plates manufactured out of specification can result in a bone cut in excess 00800"(2.8448mm) |
| 2026-02-10 | Orthalign, Inc | Drill plates manufactured out of specification can result in a bone cut in excess 00800"(2.8448mm) |
| 2026-02-02 | Howmedica Osteonics Corp. | A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw |
| 2026-01-02 | Alphatec Spine, Inc. | Due a design issue where the navigated array connection geometry is incorrect. |