AIA-PACK TgAb
TOSOH CORPORATION · Model N/A · UDI-DI 14560189210490
10
510(k) clearances
310
Adverse events
18
Recalls
34
Facilities
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2025-11-21 | Roche Diagnostics Operations, Inc. | Issues identified: 1) Results vary by Instrument Platform; 2) Calibrator Lot-to-Lot Variability. Issues may lead to delayed or incorrect diagnosis, as well as delay of treatment or |
| 2025-04-24 | Mckesson Medical-Surgical Inc. Corporate Office | transit delays of certain cold chain products due to severe thunderstorms which may caused the products to be stored outside the recommended labeled storage conditions. This failur |
| 2025-04-24 | Mckesson Medical-Surgical Inc. Corporate Office | transit delays of certain cold chain products due to severe thunderstorms which may caused the products to be stored outside the recommended labeled storage conditions. This failur |
| 2024-08-21 | Siemens Healthcare Diagnostics, Inc. | Positive bias in patient results at and below the cut-off of 60 U/mL with the lots listed in the table below when compared to alternate lots on the Atellica IM analyzer. |
| 2024-08-21 | Siemens Healthcare Diagnostics, Inc. | Positive bias in patient results at and below the cut-off of 60 U/mL with the lots listed in the table below when compared to alternate lots on the Atellica IM analyzer. |
| 2024-04-23 | Siemens Healthcare Diagnostics, Inc. | The potential for falsely elevated patient sample results when using impacted lots (785, 790, 791, and 793) of IMMULITE 2000 Anti-TG Ab. |
| 2024-04-23 | Siemens Healthcare Diagnostics, Inc. | The potential for falsely elevated patient sample results when using impacted lots (782, 783, 785, 786, 787, 788, 789, 790, 791, 792 and 793), of IMMULITE 2000 Anti-TG Ab. |
| 2023-04-24 | Abbott Ireland Limited | Due to a manufacturing fill volume error, Alinity i Anti-TPO Reagent contains a marginally lower fill volume than detailed in the Kit Contents section of the Instructions for Use ( |
| 2022-08-11 | Siemens Healthcare Diagnostics, Inc. | An average negative bias of -23% with IMMULITE 2000/IMMULITE 2000 XPi Thyroid Stimulating Immunoglobulins (TSI) when compared to other in-date kit lots, there is a potential for er |
| 2020-08-20 | Roche Diagnostics Operations, Inc. | Recent internal verification studies found the current RF interference claim to no longer be met for the assay on the cobas e 411, 601, and 602 analyzers. Therefore, the claim has |