FDA Device Intelligence
HomeMICHIGAN CRITICAL CARE CONSULTANTS, INC. › MC3 Nautilus ECMO Oxygenator

MC3 Nautilus ECMO Oxygenator

MICHIGAN CRITICAL CARE CONSULTANTS, INC. · Model 48145E · UDI-DI 10854916006901

BYS Prescription
3
510(k) clearances
177
Adverse events (5 deaths)
2
Recalls
7
Facilities

Description

Nautilus ECMO oxygenators are diffusion membrane oxygenators used in extracorporeal life support procedures to oxygenate blood, remove carbon dioxide and regulate blood temperature. Blood enters the device and passes through both the heat exchange membrane, where temperature is adjusted, and the gas transfer membrane, where oxygen is added and carbon dioxide is removed.

Recalls

DateFirmReason
2023-10-11Mc3 IncMay have increased water path restriction within the heat exchanger, leading to degraded heat exchange performance during use. The reduction of water flow and reduced heat exchange
2023-10-11Mc3 IncMay have increased water path restriction within the heat exchanger, leading to degraded heat exchange performance during use. The reduction of water flow and reduced heat exchange

Query this across every FDA database — by API.

Linked GUDID · 510(k)/PMA · MAUDE · recalls · establishments. Free tier, no card required.

Get your free API keyRead the docs
Data compiled from public U.S. FDA sources (GUDID, 510(k)/PMA, MAUDE, recalls, establishment registration). Adverse-event reports are unverified and do not establish causation. Informational only — not medical, clinical, or regulatory advice.