FDA Device Intelligence
HomeBECKMAN COULTER IRELAND INC. › iQ200

iQ200

BECKMAN COULTER IRELAND INC. · Model Sprint Pro (NA)/SPRINT w load/unload (NA)/SPRINT Pro w load/unload (NA)/Sprint P · UDI-DI 10837461001256

KQOLKM Prescription
1
510(k) clearances
1,089
Adverse events (1 deaths)
33
Recalls
65
Facilities

Description

"iQ200 Sprint 2008 Tested; 800-3053 iQ®200 Sprint Pro (NA)" "iQ200 Sprint 2008 Tested; 800-3081 iQ200 SPRINT w load/unload (NA);" "iQ200 Sprint 2008 Tested; 800-3082 iQ200 SPRINT Pro w load/unload (NA); 800-3080 IRICELL 3000 PRO w load/unload; 800-3048 IRICELL 3000 Pro (NA)" "iQ200 Sprint 2008 Tested; 800-3933 iQ®200 Sprint Plus US (2008 Version) (NA); 800-7713 IRICELL 3000 Plus (NA);800-3079 IRICELL 3000 Plus w/ Load Unload (NA)"

Recalls

DateFirmReason
2023-04-05Beckman Coulter, Inc.Beckman Coulter has become aware of an intermittent issue in which the optional flag Previous Sample Had Sperm was enabled but not displayed so that a carryover event could have be
2023-04-05Beckman Coulter, Inc.Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have
2023-04-05Beckman Coulter, Inc.Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have
2023-04-05Beckman Coulter, Inc.Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have
2021-10-18Roche Diagnostics Operations, Inc.A potential risk for false negative nitrite results exists when endogenous creatinine levels are 15,000 mg/L and above. No interference of Nitrite results was observed at creatinin
2020-03-30Beckman Coulter Inc.A defect in the SATA power adapter cable within the instrument s computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or me
2020-03-30Beckman Coulter Inc.A defect in the SATA power adapter cable within the instruments computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or mel
2020-02-17Beckman Coulter, Inc.The Leukocyte and Compensation pads were switched on some vials of lot number 7212269M of the iChem VELOCITY Strips.
2020-01-21Sysmex America, Inc.When excessive amounts of urine spills on the rack transport track and reaches the electrical components of the CV-11 conveyor, there is a potential for smoke and/or sparks which
2019-11-07Beckman Coulter Inc.Beckman Coulter has become aware of an issue affecting Specific Gravity chemistry settings for the iChemVELOCITY Urine Chemistry System that prevents the trigger of a chemistry sys

Query this across every FDA database — by API.

Linked GUDID · 510(k)/PMA · MAUDE · recalls · establishments. Free tier, no card required.

Get your free API keyRead the docs
Data compiled from public U.S. FDA sources (GUDID, 510(k)/PMA, MAUDE, recalls, establishment registration). Adverse-event reports are unverified and do not establish causation. Informational only — not medical, clinical, or regulatory advice.