Home › ESSEX INDUSTRIES, INC. › Backpack Medical Oxygen System, BMOS

Backpack Medical Oxygen System, BMOS

ESSEX INDUSTRIES, INC. · Model 50C-0089-4 · UDI-DI 10815363020056

BYJ
1
510(k) clearances
606
Adverse events (80 deaths)
6
Recalls
3
Facilities

Description

The Backpack Medical Oxygen System (BMOS) is a low pressure, portable 2.0 liter liquid oxygen storage and gaseous delivery system, capable of supplying up to 1720 gaseous liters of oxygen at a maximum flow rate of 15 liters per minute at 50 psig. A built-in flow control valve delivers oxygen to a single patient at flow rates from 0.5 to 15 liters per minute. A secondary accessory port can supply 50 psi of oxygen to other medical devices or to an oxygen regulator for a second patient. The BMOS has been successfully tested to work in tandem with commonly used ventilators. Weighing just 19 pounds when full, the BMOS can be carried or worn by parachutists and ground support personnel. The BMOS requires no external power and uses just six AA-batteries for the quantity indicator. The system has been tested to MIL-STD-810.

Recalls

DateFirmReason
2022-10-10Caire, Inc.An audit discovered some inconsistencies in weld penetration on the longitudinal weld seam on the inner bottle of the device on some of the units.
2022-10-10Caire, Inc.An audit discovered some inconsistencies in weld penetration on the longitudinal weld seam on the inner bottle of the device on some of the units.
2022-10-10Caire, Inc.An audit discovered some inconsistencies in weld penetration on the longitudinal weld seam on the inner bottle of the device on some of the units.
2022-10-10Caire, Inc.An audit discovered some inconsistencies in weld penetration on the longitudinal weld seam on the inner bottle of the device on some of the units.
2007-10-23Caire, Inc.Moisture ingress into the electronic liquid level indicator may cause inaccurate liquid oxygen level readings. The oxygen could run out without the patient knowledge.
2006-10-04Precision Medical, Inc.Component assembly; The fill chuck is not assembled properly which may allow the device to come apart over time releasing its contents.

Query this across every FDA database — by API.

Linked GUDID · 510(k)/PMA · MAUDE · recalls · establishments. Free tier, no card required.

Get your free API keyRead the docs
Data compiled from public U.S. FDA sources (GUDID, 510(k)/PMA, MAUDE, recalls, establishment registration). Adverse-event reports are unverified and do not establish causation. Informational only — not medical, clinical, or regulatory advice.