The Rapid Response® U121S Urine Analyzer is intended for use in conjunction with the Rapid Response® Urinalysis Reagent Strips for the semi-quantitative detection of analytes in human urine. The measurement can be used in general evaluation of health, and aids in the diagnosis and monitoring of metabolic or systemic diseases that affect kidney function, endocrine disorders and diseases or disorders of the urinary tract. The instrument is intended for professional, in vitro diagnostic use only.

Recalls

DateFirmReason
2026-08-12Beckman Coulter, Inc.The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replace
2026-08-12Beckman Coulter, Inc.The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replace
2023-04-05Beckman Coulter, Inc.Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have
2023-04-05Beckman Coulter, Inc.Beckman Coulter has become aware of an intermittent issue in which the optional flag Previous Sample Had Sperm was enabled but not displayed so that a carryover event could have be
2023-04-05Beckman Coulter, Inc.Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have
2023-04-05Beckman Coulter, Inc.Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have
2021-10-18Roche Diagnostics Operations, Inc.A potential risk for false negative nitrite results exists when endogenous creatinine levels are 15,000 mg/L and above. No interference of Nitrite results was observed at creatinin
2020-03-30Beckman Coulter Inc.A defect in the SATA power adapter cable within the instrument s computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or me
2020-03-30Beckman Coulter Inc.A defect in the SATA power adapter cable within the instruments computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or mel
2020-02-17Beckman Coulter, Inc.The Leukocyte and Compensation pads were switched on some vials of lot number 7212269M of the iChem VELOCITY Strips.