FDA Device Intelligence
HomeLEICA BIOSYSTEMS MELBOURNE PTY LTD › BOND-MAX Processing Module

BOND-MAX Processing Module

LEICA BIOSYSTEMS MELBOURNE PTY LTD · Model BOND-MAX · UDI-DI 09349458005689

KPA Prescription
10
510(k) clearances
1,446
Adverse events (1 deaths)
30
Recalls
83
Facilities

Description

The BOND system automates clinical protocols for immunostaining of pathology specimens mounted on microscope slides. Microscope slides subsequently undergo interpretation by a qualified healthcare professional to aid diagnosis.

Recalls

DateFirmReason
2024-04-02Agilent TechnologiesThere is the potential for the front plexiglass door of the automated slide processing system to break causing potential injury.
2022-06-01Ventana Medical Systems IncThere is a potential for fluid leak inside the Ventana HE 600 instrument which could result in electrical short and fire
2022-01-20Ventana Medical Systems IncPotential for Fluid Leak inside a staining system that could cause an electrical short circuit of the General Purpose Input/ Output Board's J1 Connector that could result in smoke
2021-04-16Hardy DiagnosticsInstruments were assembled and serviced using a defective lot of tubing, which is part of the pump mechanism within the unit.
2021-02-25Biocare Medical, LLCTheir is a potential that the automated staining instrument with software version 3.5.3.1 may move with random speeds in random direction across its range of motion. This could re
2020-08-31Roche Diagnostics Operations, Inc.Potential for Fluid leaks within the instruments, creating an electrical short inside the power socket, which generates heat, smoke, and eventually burning and melting of the elect
2020-08-31Roche Diagnostics Operations, Inc.Potential for Fluid leaks within the instruments, creating an electrical short inside the power socket, which generates heat, smoke, and eventually burning and melting of the elect
2019-11-05Dako Denmark A/SThe Autostainer instrument may not have a syringe tray installed which could result in a small volume of excess buffer being applied to several slides. In addition to the syringe t
2019-10-04Ventana Medical Systems IncThe firm became aware of a cleaning solution leaking issue from the middle staining module of the Ventana HE 600 instrument, used for histologic section samples in anatomic patholo
2018-09-17Leica Microsystems, Inc.The power supply unit for the printer used with the device can potentially overheat or cause a fire hazard because it could contain an improperly-mixed phosphorus compound.