FDA Device Intelligence
HomeHDX WILL CORP. › STAR-X

STAR-X

HDX WILL CORP. · Model STAR-X · UDI-DI 08800005200151

EHD Prescription
1
510(k) clearances
521
Adverse events
18
Recalls
63
Facilities

Description

Star-X is intra-oral X-ray system for dental diagnosis that can radiate X-ray images by irradiating X-rays generated by high voltage to an anode in an X-ray tube and reacting with the receptor (film, sensor).Star-X is a dental X-ray equipment (extra-oral source system) intended for use by qualified dentist or dental technician for both adult and pediatric patient for producing diagnostic dental radiographs for treatment of diseases of the teeth, jaw, and other oral structures using intra-oral image receptors.

Recalls

DateFirmReason
2025-06-19DIGIMED CO., LTDSome DIOX-602 units currently in use within the United States do not meet the minimum SSD requirement of 18 cm.
2017-07-14Panoramic Rental Corp.The lift motor used to raise and lower the overhead chassis will wear if not properly maintained according to the maintenance schedule.
2016-08-19Panoramic Rental Corp.Panoramic Corporation is performing a voluntary Medical Device Correction on certain PC-1000 X-ray systems due to the potential for a safety nut to be ineffective, which could resu
2016-07-18Aribex IncFirm received customer complaints where end users reported that Nomad Pro and Pro 2 handsets had appeared to "overheat", "melt", "smoke" and/or "ignite." Investigation found that
2016-07-18Aribex IncFirm received customer complaints where end users reported that Nomad Pro and Pro 2 handsets had appeared to "overheat", "melt", "smoke" and/or "ignite." Investigation found that
2015-06-04Panoramic Rental Corp.Jackscrew nut part failure potentially leading to device malfunction and patient/operator injury.
2014-08-12Gendex CorpTwo (2) failures of the spring link components within the articulated arm assembly. The spring links broke during normal assembly process for the articulated arm during first cycle
2014-01-24Carestream Health Inc.Carestream received problem reports relating to apparent early failure of Kodak 2X00 Intraoral X-ray device arms leading to arm breakage.
2014-01-24Carestream Health Inc.Carestream received problem reports relating to apparent early failure of Kodak 2X00 Intraoral X-ray device arms leading to arm breakage.
2014-01-14Instrumentarium Dental, PaloDEx Group OyInstrumentarium Dental has delivered a number of FOCUS 3.0 Intraoral X-ray units in November 2013 with a potential flaw in the mounting hardware. A number of wall mount units were

Query this across every FDA database — by API.

Linked GUDID · 510(k)/PMA · MAUDE · recalls · establishments. Free tier, no card required.

Get your free API keyRead the docs
Data compiled from public U.S. FDA sources (GUDID, 510(k)/PMA, MAUDE, recalls, establishment registration). Adverse-event reports are unverified and do not establish causation. Informational only — not medical, clinical, or regulatory advice.