Adherent Bag
Scinus Cell Expansion Netherlands B.V. · Model ADH-01-01 · UDI-DI 08720299213616
Description
For the adherent cell culture inside the SCINUS the Adherent Bag is used. This Adherent Bag is intended exclusively for use with the SCINUS. The SCINUS is a closed, controlled bioreactor system for automated cell cultivation. Combined with the Adherent Bag, it allows large-fold cell expansion without passing the cells due to its unique expansion technology. The SCINUS is operated by a controller, which is accessible through an integrated graphical user interface (GUI). The unique combination of the rocker, roller, and Adherent Bag ensures a homogenous environment with minimal levels of shear forces. The roller on top of the rocker enables the expansion of the available culture volume and surface area. The pH and dissolved oxygen (DO) levels are measured by integrated sensors in the Adherent Bag and are controlled with the controller software. The carboxygenator ensures that the set points are maintained within the desired bandwidths. For the culture of adherent cells, the Adherent Bag must be installed in the SCINUS. The Adherent Bag contains the culture bag and the corresponding weldable tubing set. One of the unique features of the Adherent Bag is the gradual expansion of the cell culture volume. The volume can be increased from a minimal volume of 0.1 L to 1.4 L to accommodate the growing cell population.
Query this across every FDA database — by API.
Linked GUDID · 510(k)/PMA · MAUDE · recalls · establishments. Free tier, no card required.
Get your free API keyRead the docs