FDA Device Intelligence
HomeVarex Imaging Nederland B.V. › Claymount

Claymount

Varex Imaging Nederland B.V. · Model 50126-27-OEM · UDI-DI 08718755000511

IZW Prescription
10
510(k) clearances
9
Adverse events
3
Recalls
17
Facilities

Description

x-ray beam-limiting device, to restrict the dimensions of a diagnostic x-ray field by limiting the size of the primary x-ray beam.(Collimator) This is a device out of the Optica 40 series

Recalls

DateFirmReason
2011-10-28Shimadzu Medical SystemsShimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standard
2008-02-07Philips Medical Systems North America Co. PhillipsThe collimator could fall from the x-ray stand, due to loosening screws, and strike a patient or operator underneath.
2007-03-15Philips Medical Systems North America Co. PhillipsThe Collimator field-defining light beam does not meet the light field contrast ratio requirement of 4:1. System is non-compliant with performance standard (1020.31 (d)(2)(iii) 21