FDA Device Intelligence
Home › INSTRUMENTATION LABORATORY COMPANY › GEM Premier 5000 PAK BG COOX TBILI (300 TEST)

GEM Premier 5000 PAK BG COOX TBILI (300 TEST)

INSTRUMENTATION LABORATORY COMPANY · Model 00055430005XC · UDI-DI 08426951027248

CHL Prescription
1
510(k) clearances
2,419
Adverse events (49 deaths)
185
Recalls
34
Facilities

Description

GEM 5000 BG COOX TBILI 300 Test PAK 31 DAY (pH, pCO2, pO2, Hct, tHb, O2Hb, COHb, HHb, MetHb, sO2, tBili Cartridge)

Recalls

DateFirmReason
2026-08-18Radiometer Medical ApSIn vitro diagnostic analyzer had a risk of potentially inaccurate results requiring changes to instructions for use and: 1) Sample types, including whole blood samples no longer su
2026-05-21Instrumentation LaboratoryThe active ingredient concentration in the lysing bag solution of specific GEM Premier 5000 lots did not meet established manufacturing specifications. The malfunction could result
2026-05-21Instrumentation LaboratoryThe active ingredient concentration in the lysing bag solution of specific GEM Premier 7000 with iQM3 PAK lots did not meet established manufacturing specifications. The malfunctio
2026-05-08Instrumentation LaboratoryPotential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
2026-05-08Instrumentation LaboratoryPotential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
2026-05-08Instrumentation LaboratoryPotential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
2026-05-08Instrumentation LaboratoryPotential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
2026-05-08Instrumentation LaboratoryPotential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
2026-05-08Instrumentation LaboratoryPotential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
2026-04-02Siemens Healthcare Diagnostics IncUse of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary