The device is a washer disinfector dedicated to the treatment of cart and trolley. It offers complete cleaning and thermal disinfection (intermediate level) treatment.
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2025-11-19 | Steris Corporation | Wire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of s |
| 2025-11-19 | Steris Corporation | Wire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of s |
| 2025-08-29 | Getinge Disinfection Ab | Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not tri |
| 2025-04-02 | Getinge Disinfection Ab | The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and t |
| 2025-04-02 | Getinge Disinfection Ab | The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and t |
| 2023-04-10 | Steris Corporation | In the remote occurrence in which the electrical contactor component present in the drying chambers of the Reliance Vision Multi-Chamber Washer/Disinfector malfunctions, the heatin |
| 2022-04-22 | Steris Corporation | The firm identified that when the electrical contactor component in the drying chamber of the Washer/Disinfector malfunctions, the heating elements in the drying chamber could over |
| 2022-04-13 | Getinge Usa Sales Inc | Verification of installation not documented or may be incomplete, could result in harm to the user such as skin burns, smoke inhalation, electric shock, or a crush or pinch injury |
| 2022-04-13 | Getinge Usa Sales Inc | Verification of installation not documented or may be incomplete, could result in harm to the user such as skin burns, smoke inhalation, electric shock, or a crush or pinch injury |
| 2022-04-01 | Getinge Usa Sales Inc | Inadequate documentation verifying whether device installation has been executed in accordance with the manufacturer's requirements. This can lead to installation failure modes and |