The Lipase NG assay is an in vitro diagnostic test used for the determination on AU480 and AU680 automatic analyzers of the lipase activity in serum and plasma by kinetic colorimetric method. The assay is intended for professional use only.

Recalls

DateFirmReason
2023-09-12Ortho-Clinical Diagnostics, INc.Products 1668409 and 8297749 may contain VITROS Lipase Slides with an abnormality which could cause -46% to +86% bias in lipase results depending on how the slides are used. Falsel
2020-06-26RANDOX LABORATORIES, LTD.Randox have confirmed imprecision of quality control and patient samples when using specific Randox Lipase buffer batches.
2020-06-26RANDOX LABORATORIES, LTD.Randox have confirmed imprecision of quality control and patient samples when using specific Randox Lipase R2 Substrate batches.
2020-06-26RANDOX LABORATORIES, LTD.Randox have confirmed imprecision of quality control and patient samples when using specific Randox Lipase reagent batches.
2019-12-20Siemens Healthcare Diagnostics, Inc.Firm has received customer complaints regarding failed calibrations and increased imprecision of quality control and patient samples when using this product.
2019-02-06Medica CorporationWash1 solution is ineffective in cleaning dispensing probe of TRIG and LDL reagents that interfere with the Lipase (LIP) assay on the EasyRA clinical chemistry analyzer.
2018-10-30Horiba Instruments IncorporatedPotential for chemical analyzer to report abnormal results due to an interference with triglycerides CP and Cholesterol CP. An over estimation of the lipase concentration could b
2018-09-07Siemens Healthcare Diagnostics, Inc.There is lot to lot variability for QC and patient results when using the identified reagent cartridge lots.
2018-09-07Siemens Healthcare Diagnostics, Inc.There is lot to lot variability for QC and patient results when using the identified reagent cartridge lots.
2017-12-14Siemens Healthcare Diagnostics, Inc.Calibration failure - confirmed a positive bias with LIPL lot FB8109 when compared to other Dimension LIPL lots. Lot FB8109 is not meeting Siemens internal specifications for lot-t