Lipase NG
SENTINEL CH. SPA · Model B82715 · UDI-DI 08058056681782
CHI
Prescription
10
510(k) clearances
98
Adverse events
22
Recalls
29
Facilities
Description
The Lipase NG assay is an in vitro diagnostic test used for the determination on AU480 and AU680 automatic analyzers of the lipase activity in serum and plasma by kinetic colorimetric method. The assay is intended for professional use only.
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2023-09-12 | Ortho-Clinical Diagnostics, INc. | Products 1668409 and 8297749 may contain VITROS Lipase Slides with an abnormality which could cause -46% to +86% bias in lipase results depending on how the slides are used. Falsel |
| 2020-06-26 | RANDOX LABORATORIES, LTD. | Randox have confirmed imprecision of quality control and patient samples when using specific Randox Lipase buffer batches. |
| 2020-06-26 | RANDOX LABORATORIES, LTD. | Randox have confirmed imprecision of quality control and patient samples when using specific Randox Lipase R2 Substrate batches. |
| 2020-06-26 | RANDOX LABORATORIES, LTD. | Randox have confirmed imprecision of quality control and patient samples when using specific Randox Lipase reagent batches. |
| 2019-12-20 | Siemens Healthcare Diagnostics, Inc. | Firm has received customer complaints regarding failed calibrations and increased imprecision of quality control and patient samples when using this product. |
| 2019-02-06 | Medica Corporation | Wash1 solution is ineffective in cleaning dispensing probe of TRIG and LDL reagents that interfere with the Lipase (LIP) assay on the EasyRA clinical chemistry analyzer. |
| 2018-10-30 | Horiba Instruments Incorporated | Potential for chemical analyzer to report abnormal results due to an interference with triglycerides CP and Cholesterol CP. An over estimation of the lipase concentration could b |
| 2018-09-07 | Siemens Healthcare Diagnostics, Inc. | There is lot to lot variability for QC and patient results when using the identified reagent cartridge lots. |
| 2018-09-07 | Siemens Healthcare Diagnostics, Inc. | There is lot to lot variability for QC and patient results when using the identified reagent cartridge lots. |
| 2017-12-14 | Siemens Healthcare Diagnostics, Inc. | Calibration failure - confirmed a positive bias with LIPL lot FB8109 when compared to other Dimension LIPL lots. Lot FB8109 is not meeting Siemens internal specifications for lot-t |