Home › SENTINEL CH. SPA › Ammonia Ultra
Ammonia Ultra
SENTINEL CH. SPA · Model 17660 · UDI-DI 08058056680181
JIF
Prescription
1
510(k) clearances
50
Adverse events (1 deaths)
11
Recalls
18
Facilities
Description
The Ammonia Ultra assay is intended for the in vitro quantitative enzymatic determination of ammonia in human plasma using an automated analyzer. Measurements of ammonia are used to aid in the diagnosis of patients suspected of having hyperammonemia. For laboratory professional use only.
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2020-07-30 | RANDOX LABORATORIES, LTD. | Randox Ammonia reagents are being recalled from the field due to a positive bias of up to 140mol/l being observed on patient samples. |
| 2020-07-30 | RANDOX LABORATORIES, LTD. | Randox Ammonia reagents are being recalled from the field due to a positive bias of up to 140mol/l being observed on patient samples. |
| 2017-06-29 | Siemens Healthcare Diagnostics, Inc. | Various assays used to measure different chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridine causing falsely ele |
| 2017-06-29 | Siemens Healthcare Diagnostics, Inc. | Various assays used to measure different chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridine causing falsely ele |
| 2017-06-29 | Siemens Healthcare Diagnostics, Inc. | Various assays used to measure different chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridine causing falsely ele |
| 2016-12-16 | Siemens Healthcare Diagnostics, Inc. | Siemens Healthcare Diagnostics has determined that Dimension AMM (DF119) flex reagent cartridge lots EB7180, FB7152, BA7194, EA7223, BA7250 do not meet the 60-day calibration inter |
| 2016-12-16 | Siemens Healthcare Diagnostics, Inc. | Siemens Healthcare Diagnostics has determined that Dimension Vista AMM (K3119) Flex reagent cartridge flex lots 16187BE, 16225BB, 16265AB do not meet the 60-day calibration interva |
| 2016-11-29 | Sekisui Diagnostics P.E.I. Inc. | Certain lots are showing an atypical decrease in optical density (OD) over time, which may result in an impact to performance at the high end of the linear range. |
| 2015-09-22 | Sentinel CH SpA | Results for ammonia are below the linear range of the assay for samples that have been collected in sodium or lithium heparin tubes. |
| 2009-03-26 | Beckman Coulter Inc. | The Ammonia Reagent lot M808278 may fail calibration or cause a low shift in Quality Control recovery on some SYNCHRON Systems. |