FDA Device Intelligence
HomeEL.EN. SPA › DEKA LIPOAI

DEKA LIPOAI

EL.EN. SPA · Model M131A1 · UDI-DI 08057017760580

GEX Prescription
1
510(k) clearances
44,032
Adverse events (107 deaths)
201
Recalls
485
Facilities

Description

The LipoAI is medical laser equipped with a short pulse 1444 nm Nd:YAG laser source. The device delivers the laser energy through a 600 um optical fiber. The fiber is protected by a small stainless-steel cannula (max. 1.4mm) and is inserted in the tissue through a small aperture in the skin.

Recalls

DateFirmReason
2026-08-06Lutronic CorporationUtilizing higher fluences at 1064nm wavelength with current cryogen parameters may result in a spark/loud noise during treatment, potentially resulting in patient burn.
2026-05-05IPG Medical CorporationSoftware anomaly that causes a false display of error code 5018.
2026-02-27Olympus Corporation of the AmericasFollowing a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperabl
2026-02-27Olympus Corporation of the AmericasFollowing a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperabl
2026-01-27Lutronic CorporationReports of devices sparking/popping and potentially burning patients.
2025-12-15LEASEIR TECHNOLOGIES SLUThe console label for affected devices is missing the "DANGER" symbol.
2025-07-22LUMENIS, LTD.The potential for unsterilized product within finished product labeled as sterile.
2025-05-06Olympus Corporation of the AmericasPotential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or
2025-05-06Olympus Corporation of the AmericasPotential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or
2025-05-06Olympus Corporation of the AmericasPotential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or