FDA Device Intelligence
Home › Dent4You AG › HIP Ultra Cleaning Solution

HIP Ultra Cleaning Solution

Dent4You AG · Model CS-HIPC-U · UDI-DI 07640397420390

LDS
10
510(k) clearances
95
Adverse events
11
Recalls
11
Facilities

Description

HIP Ultra Cleaning Solution with Instrument Protection. For the Hydrim C61 and Eschmann Little Sister.

Recalls

DateFirmReason
2018-04-09Cenorin, LLCThere is a supplier quality issue with the terminal blocks which may cause the terminal block to overheat, char and/or burn.
2016-04-22Natus Medical IncorporatedThe replacement HEPA Filters have been manufactured with sealant that may not be properly cured. Improperly cured sealant can emit a toxic gas when heated, and can drip down during
2014-04-24Cenorin, LLCCenorin LLC is voluntarily recalling HLD Systems Model 610 Washer/High Level Disinfector due to a potential risk of the system not injecting the proper amount of cleaning fluid dur
2013-01-24Cenorin, LLCThe temperature sensor/control system in the HLD Systems Model 610 may provide a false temperature reading. This caused the water to remain at ambient temperature throughout the pa
2010-06-07CenorinThe water level sensor switch in the Medical Device Cleaning and High Level Disinfection Systems (Washer/Pasteurizers) is sensitive to externally induced vibration noise, and it pr
2010-06-07CenorinThe water level sensor in the Medical Device Cleaning and High Level Disinfection (Washer/Pasteurizer) is sensitive to externally induced vibration noise, and it produces a false
2010-06-07CenorinThe water level sensor switch in the Medical Device Cleaning and High Level Disinfection Systems (Washer/Pasteurizers) is sensitive to externally induced vibration noise, and it pr
2010-06-07CenorinThe water level sensor switch in the Medical Device Cleaning and High Level Disinfection Systems (Washer/Pasteurizers) is sensitive to externally induced vibration noise, and it pr
2006-12-21CenorinPotential for user to receive an electrical shock.
2006-12-21CenorinPotential for user to receive an electrical shock.

Query this across every FDA database — by API.

Linked GUDID · 510(k)/PMA · MAUDE · recalls · establishments. Free tier, no card required.

Get your free API keyRead the docs
Data compiled from public U.S. FDA sources (GUDID, 510(k)/PMA, MAUDE, recalls, establishment registration). Adverse-event reports are unverified and do not establish causation. Informational only — not medical, clinical, or regulatory advice.