Home › Roche Diagnostics GmbH › Iron Gen.2
Iron Gen.2
Roche Diagnostics GmbH · Model 08057931190 · UDI-DI 07613336121344
JIY
Prescription
10
510(k) clearances
55
Adverse events
19
Recalls
36
Facilities
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2025-11-03 | GET TESTED INTERNATIONAL AB | Distribution without premarket approval/clearance. |
| 2024-03-13 | Siemens Healthcare Diagnostics, Inc. | Potential for falsely elevated Chol_2, LDLC, and Trig_2 results on the Atellica CH and Atellica CI analyzers when the previous result in the cuvette was Iron3. Results in a positiv |
| 2023-09-29 | Sentinel CH SpA | A high recovery of Iron assay has been observed at some customer sites. Internal studies confirmed the positive shift of Quality Control and patient samples results, up to 15% for |
| 2023-09-29 | Sentinel CH SpA | A high recovery of Iron assay has been observed at some customer sites. Internal studies confirmed the positive shift of Quality Control and patient samples results, up to 15% for |
| 2022-06-28 | Siemens Healthcare Diagnostics, Inc. | Falsely Elevated Atellica CH Microalbumin_2 (ALB_2) Results due to Reagent Carryover from the Iron_2 Assay. This defect occurs only when a microalbumin test is run immediately aft |
| 2016-08-25 | Siemens Healthcare Diagnostics, Inc. | Siemens Healthcare Diagnostics has become aware of deferoxamine interference with Dimension and Dimension Vista IRON assays at a different concentration than listed in the Instruct |
| 2016-08-25 | Siemens Healthcare Diagnostics, Inc. | Siemens Healthcare Diagnostics has become aware of deferoxamine interference with Dimension and Dimension Vista IRON assays at a different concentration than listed in the Instruct |
| 2016-07-20 | Sterling Diagnostics, Inc. | Sterling Diagnostics, Inc. initiated a recall on 7/20/2016 for multiple diagnostic kits due to a recision of the previously extended expiration dates for the kits. |
| 2013-06-12 | Ortho-Clinical Diagnostics | Ortho Clinical Diagnostics is recalling Slide Dispense cartridges used with some VITROS Chemistry Product Slides due to slide dispense issues. |
| 2012-10-08 | Siemens Healthcare Diagnostics, Inc. | Siemens Healthcare Diagnostics has confirmed the potential for falsely elevated results with IRON Flex(R) reagent cartridges when tests are processed from a well set contaminated w |