FDA Device Intelligence
HomeRoche Diagnostics GmbH › Iron Gen.2

Iron Gen.2

Roche Diagnostics GmbH · Model 08057931190 · UDI-DI 07613336121344

JIY Prescription
10
510(k) clearances
55
Adverse events
19
Recalls
36
Facilities

Recalls

DateFirmReason
2025-11-03GET TESTED INTERNATIONAL ABDistribution without premarket approval/clearance.
2024-03-13Siemens Healthcare Diagnostics, Inc.Potential for falsely elevated Chol_2, LDLC, and Trig_2 results on the Atellica CH and Atellica CI analyzers when the previous result in the cuvette was Iron3. Results in a positiv
2023-09-29Sentinel CH SpAA high recovery of Iron assay has been observed at some customer sites. Internal studies confirmed the positive shift of Quality Control and patient samples results, up to 15% for
2023-09-29Sentinel CH SpAA high recovery of Iron assay has been observed at some customer sites. Internal studies confirmed the positive shift of Quality Control and patient samples results, up to 15% for
2022-06-28Siemens Healthcare Diagnostics, Inc.Falsely Elevated Atellica CH Microalbumin_2 (ALB_2) Results due to Reagent Carryover from the Iron_2 Assay. This defect occurs only when a microalbumin test is run immediately aft
2016-08-25Siemens Healthcare Diagnostics, Inc.Siemens Healthcare Diagnostics has become aware of deferoxamine interference with Dimension and Dimension Vista IRON assays at a different concentration than listed in the Instruct
2016-08-25Siemens Healthcare Diagnostics, Inc.Siemens Healthcare Diagnostics has become aware of deferoxamine interference with Dimension and Dimension Vista IRON assays at a different concentration than listed in the Instruct
2016-07-20Sterling Diagnostics, Inc.Sterling Diagnostics, Inc. initiated a recall on 7/20/2016 for multiple diagnostic kits due to a recision of the previously extended expiration dates for the kits.
2013-06-12Ortho-Clinical DiagnosticsOrtho Clinical Diagnostics is recalling Slide Dispense cartridges used with some VITROS Chemistry Product Slides due to slide dispense issues.
2012-10-08Siemens Healthcare Diagnostics, Inc.Siemens Healthcare Diagnostics has confirmed the potential for falsely elevated results with IRON Flex(R) reagent cartridges when tests are processed from a well set contaminated w

Query this across every FDA database — by API.

Linked GUDID · 510(k)/PMA · MAUDE · recalls · establishments. Free tier, no card required.

Get your free API keyRead the docs