FDA Device Intelligence
HomeO-Two Medical Technologies Inc › EASY Grip

EASY Grip

O-Two Medical Technologies Inc · Model 01BM3520 · UDI-DI 07540195000285

BTM Prescription
10
510(k) clearances
1,891
Adverse events (164 deaths)
97
Recalls
74
Facilities

Description

The disposable infant EASY GRIP® Bag Valve Mask Resuscitator is a reusable device for manual ventilation. It is designed for use in various clinical settings to provide respiratory support to the respiratory distressed or non-breathing INFANT patient. When used with a supplemental oxygen supply, the device may provide 100% oxygen on every squeeze of the bag. The disposable Infant EASY GRIP resuscitator comes standard with a Pressure Relief Valve set at 40 cm H2O which can be overridden by the operator.

Recalls

DateFirmReason
2026-05-08Datex-Ohmeda IncBlender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration se
2025-07-09Ambu Inc.Potential for the manometer port being blocked rendering the manometer non-functional.
2025-05-01SunMed Holdings, LLCAffected lots were manufactured with B/V Filter incorrectly attached to the wrong port (patient port instead of the exhalation port). If not noticed prior to patient use, there wou
2024-01-10Vyaire MedicalRisk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ven
2024-01-10Vyaire MedicalRisk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ven
2024-01-10Vyaire MedicalRisk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ven
2024-01-10Vyaire MedicalRisk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ven
2024-01-10Vyaire MedicalRisk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ven
2024-01-10Vyaire MedicalRisk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ven
2024-01-10Vyaire MedicalRisk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ven