Description

Sectra Workstation is intended for the display, manipulation and diagnostic review of medical images (e.g. DICOM, jpeg, mpeg) and other related medical information. It can show images from different sources and interface with various image storage and printing devices. Areas of application include e.g. radiology (e.g. CR, DX, CT, MR, US), mammography, cardiology and pathology. Typical users are trained medical professionals, including, but not limited to physicians, radiologists, nurses, medical technicians and assistants. Lossy compressed radiological images and DICOM “for processing” images are not intended for diagnostic review. In the United States only: - Mammographic images must be reviewed on a monitor approved by FDA for mammographic images - The Sectra Workstation is not intended for primary diagnosis of digital pathology images. Sectra UniView, the web-based model of Sectra Workstation, is primarily intended for reference viewing on both stationary and mobile platforms. It is not intended for replacing a full diagnostic workstation but can be used on stationary platforms for diagnostic review of medical images except mammograms. For diagnostic review of radiological images full image quality mode must be used. Note: The user is responsible for the quality, accuracy and adequacy of all information imported or entered into Sectra Workstation, whether videos or photographs, audio or text. Further, the user is responsible for the clinical adequacy of the viewing device used (e.g., size, resolution and contrast).

Recalls

DateFirmReason
2026-07-22Merge Healthcare, Inc.Software issue may cause measurements to be displayed inaccurately.
2026-07-22FUJIFILM Healthcare Americas CorporationSoftware issue may cause the ruler measurement to be inaccurate.
2026-06-30Merge Healthcare, Inc.Merge Healthcare has identified a potential issue that may cause thumbnails from a previously opened patient study to be displayed in the Thumbnail Bar while a different patient st
2026-06-30Merge Healthcare, Inc.Merge Healthcare has identified a potential issue that may cause thumbnails from a previously opened patient study to be displayed in the Thumbnail Bar while a different patient st
2026-06-19Boston Scientific CorporationSoftware anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more input. When a user modifies an inpu
2026-05-08GE Medical Systems SCSGE HealthCare has become aware of a context synchronization issue in AW Server 3.2 ext. 6.5. When a user selects a patient or exam in the AW Server Web Client worklist and launches
2026-03-19PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Potential safety issue involving the Mean Aneurysm Flow Amplitude (MAFA) ratio of AneurysmFlow (Interventional Tool). Philips has determined that the MAFA ratio does not provide re
2026-03-05PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Potential that mis-ordered frames in Vue Motion during dynamic cine runs may cause images frames to display out of sequence.
2026-01-30GE Medical Systems, LLCThere is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, w
2026-01-30GE Medical Systems, LLCThere is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, w