FDA Device Intelligence
HomeCellavision AB › CellaVision® Remote Review Software, Team Edition, version independent

CellaVision® Remote Review Software, Team Edition, version independent

Cellavision AB · Model CRRS-TE0099 · UDI-DI 07350040975909

JOY Prescription
1
510(k) clearances
18
Adverse events
4
Recalls
10
Facilities

Description

The CellaVision remote review software enable remote users to access data collected, stored and processed by CellaVision DM Instruments with the ability to reclassify cells and sign off slides from another location.

Recalls

DateFirmReason
2025-10-08Cellavision ABAutomated cell-locating device barcode reader may read the barcode of the previously processed slide resulting in a misattribution of diagnostic results.
2018-05-07Roche Diagnostics HematologyDiscrepant results have been reported in patients with severe microcytic anemia (e.g.iron deficiency, thalassemia) and the following parameters are affected: RBC (red blood cell co
2017-12-04Cellavision ABA software malfunction was found where WBC, RBC and PLT comments added after a slide is signed, are not sent to the LIS. This can only occur where customers process multiple slides
2016-05-02Leica Biosystems Richmond Inc.Systems have an improperly activated Windows 7 OS, even though a valid Windows 7 license was associated with each manufactured system. The system appears to be unlicensed. These