IMAGE NAVIGATION LTD · Model A-IG-0005-V1.5 · UDI-DI 07290016855012
Description
The Image-Guided Implantation system is an image-guided surgery system, designed to provide guidance of surgical instruments used in dental implantation procedures. The system implements the well-established advances in the fields of computer-aided surgery and image-guided surgery into the field of dental implantology. The IGI-system provides accurate guidance of the surgical instruments with regard to the patient’s anatomical structures and to the pre-surgical plan, based on the patient’s CT imaging. The system is an infra-red optics-based navigational system, which utilizes an infra-red camera sensor and light-emitting diodes (LEDs) to monitor objects in a defined space. The system comprises of the following components: computer processing unit, monitor, tracking camera and trackable appliances (designed to monitor the surgical hand-piece and patient position).
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2026-03-19 | Elekta, Inc. | Locking mechanisms may not properly secure to the Arc system, resulting in shifting of coordinates during procedures. |
| 2025-11-12 | Howmedica Osteonics Corp. | When the device is activated during surgery, the device software issues an error message "Instrument Face Mask (6001385000000) is not valid for this application" resulting in an in |
| 2025-09-30 | Elekta, Inc. | Due to increased risk that paint flakes from the engravings may separate from the device and there is the possibility that the flake may inadvertently enter the wound |
| 2024-10-04 | FHC, Inc. | A complaint was received (MDR 3002250546-2024-00003) prompting an investigation. The investigation revealed a schematic for a spacer tube which is too long relative to the size of |
| 2024-09-25 | Elekta Instrument AB | Elekta has become aware that Disposable Biopsy Needles can contain some microscopic debris on the inside of the biopsy needle. The material in the debris is stainless steel, same m |
| 2024-03-26 | Medtronic Navigation, Inc. | Due to a software issue, there is the potential for a missing digit, letter, or decimal in either the "distance to target" or "tip stop point" display during navigation. |
| 2024-03-15 | Elekta Instrument AB | Elekta has become aware that Disposable Biopsy Needles (911933) from one batch (837838839) can contain some microscopic debris on the inside of the biopsy needle. The material in t |
| 2024-02-26 | Navinetics Inc | Defective Skull Anchor Key. After attachment of the single use Skull Anchor Key to the patient, the user was unable to attach the CT Localizer to perform stereotactic Imaging. Atta |
| 2023-12-11 | Globus Medical, Inc. | Sterile-packed Cranial Drills may not be sterile due to a breach in the packaging and/or may have discoloration near the drill tip. |
| 2023-12-11 | Globus Medical, Inc. | Sterile-packed Cranial Drills may not be sterile due to a breach in the packaging and/or may have discoloration near the drill tip. |