FDA Device Intelligence
HomeChangsha Yingtai Instrument Co., Ltd › Yingtai

Yingtai

Changsha Yingtai Instrument Co., Ltd · Model GL21M · UDI-DI 06976301910307

JQC
10
510(k) clearances
592
Adverse events (5 deaths)
16
Recalls
164
Facilities

Description

Centrifuge is an experimental device that utilizes centrifugal force to separate substances of different densities within a liquid sample. It is widely used in fields such as biomedical science, chemical analysis, molecular biology, and industrial production. By rapidly spinning samples to create a density gradient, centrifuges provide efficient solutions for sample extraction, purification, and concentration. They can generally be categorized into high-speed or low-speed centrifuges, refrigerated or non-refrigerated centrifuges, and floor-standing or benchtop centrifuges.

Recalls

DateFirmReason
2019-02-06LW Scientific, Inc.The clear, plastic, polycarbonate hematocrit rotor may break apart during use and be ejected through the plastic housing, posing a safety hazard to the technician and potentially d
2019-02-06LW Scientific, Inc.The clear, plastic, polycarbonate hematocrit rotor may break apart during use and be ejected through the plastic housing, posing a safety hazard to the technician and potentially d
2019-02-06LW Scientific, Inc.The clear, plastic, polycarbonate hematocrit rotor may break apart during use and be ejected through the plastic housing, posing a safety hazard to the technician and potentially d
2017-10-20Terumo BCT, Inc.The patient labels included in the BMAC system procedure packs, catalog numbers 51416 and 51423, are printed with an incorrect lot number, catalog number, and expiry date; the lot
2017-10-20Terumo BCT, Inc.The patient labels included in the BMAC system procedure packs, catalog numbers 51416 and 51423, are printed with an incorrect lot number, catalog number, and expiry date; the lot
2016-02-05Beckman Coulter Inc.A defective rotor yoke on Allegra X-5 allows improper rotor seating on the hub resulting in possible rotor failure if the buckets are loaded in an imbalanced configuration.
2014-12-03Iris DiagnosticsIris International is recalling the RTH8 Rotor used in the StatSpin Express 4 Horizontal Centrifuge because the RTH8 rotor may develop cracks with use over time.
2011-07-15Global Focus Marketing & DevelopPotential for bucket and insert not seated on the rotor to cause the lid to break and buckets to be ejected.
2010-08-18StatSpin, Inc., dba Iris Sample ProcessingRotors may fail due to wire spokes breaking
2010-01-27StatSpin, Inc., dba Iris Sample ProcessingCentrifuge shield micro-switch failed due to damage, and the unit opened, ejecting pieces.

Query this across every FDA database — by API.

Linked GUDID · 510(k)/PMA · MAUDE · recalls · establishments. Free tier, no card required.

Get your free API keyRead the docs
Data compiled from public U.S. FDA sources (GUDID, 510(k)/PMA, MAUDE, recalls, establishment registration). Adverse-event reports are unverified and do not establish causation. Informational only — not medical, clinical, or regulatory advice.