The SMD-XGD/AR Galaxy Rotating Gamma System is a teletherapy device which contains 30 cobalt - 60 sources distributed on a hemispheric source carrier, which is inside a hemispheric shield. Inside the source unit is the "built-in" secondary collimator, which has 4 sizes of collimators and 1 block position. When not treating a patient, all sources are aligned with the block position of the secondary collimator, which in this case acts as a radiation shutter Treatment starts by aligning the sources to the desired collimator size, then both the source unit and the secondary collimator rotates as one unit. By rotating the cobalt -60 y-ray beams during treatment, 30 non-overlapping full 360° arcs are formed, resulting in high focal dose uniformity and small focal spot penumbra. The entire system consists of a) The y-ray treatment unit, and b) the stereotactic localization system.

Recalls

DateFirmReason
2026-02-18Elekta, Inc.Due to adapter not being properly aligned or locked in place.
2023-11-30Xcision Medical Systems, LLCWrong structures imported into the GammaPod Treatment Planning System.-system allowed an RTSS generated from a previous CT scan to be selected
2019-07-03Xcision Medical Systems, LLCOne bolt on the V motor was not fully tightened which caused a drift of .13 degrees in alignment resulting in reduced dose output.
2018-06-27Elekta Instrument ABThe margin tool in Leksell GammaPlan 11.1 systematically overestimates margin in certain areas of the volume.
2017-05-23Elekta, Inc.If the gantry module is replaced after original installation the configuration settings may be missing. The identified risk for this issue is electrical safety for technicians doin
2016-04-19Elekta, Inc.The latches of the frame adapter can be locked even if the locating pins of the frame adapter is not inserted into the corresponding holes in the coordinate frame.
2015-06-01Elekta, Inc.Memory can become corrupted when creating a fused study via drag and drop in Leksell GammaPlan 10.2.
2014-10-06Elekta, Inc.Elekta has become aware that the latches may be locked before they have been fully turned resulting in a poorly locked frame.
2011-06-23Elekta, Inc.There is a risk that the user may interpret the instructions in the manual too literally, and adjust the TMR 10 dose description to obtain an ETCPD that matches previous clinical p
2011-05-26Elekta, Inc.Several of the LMR03 actuators with bronze drive nut have failed unexpectedly, creating a potential safety hazard for the operator and patient.