Home › LIVANOVA USA, INC. › VNS Therapy® AspireSR® Model 106 Generator

VNS Therapy® AspireSR® Model 106 Generator

LIVANOVA USA, INC. · Model 106 · UDI-DI 05425025750061

LYJ Prescription
0
510(k) clearances
51,991
Adverse events (2,918 deaths)
37
Recalls
1
Facilities

Description

Implantable Pulse Generator

Recalls

DateFirmReason
2025-01-03LivaNova USA, Inc.Their is a potential that implanted pulse generators may stop delivering therapy due to an internal component issue.
2025-01-03LivaNova USA, Inc.Their is a potential that implanted pulse generators may stop delivering therapy due to an internal component issue.
2024-12-12LivaNova USA, Inc.Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result
2024-12-12LivaNova USA, Inc.Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result
2021-12-02LivaNova USA, Inc.A manufacturing issue may cause the silicone tubing wall thickness of vagus nerve stimulation leads to be less than the design specification. If the tubing becomes completely breac
2020-11-12LivaNova USA IncDuring internal testing, it was found that upon a device reset, the Generator exhibits the incorrect Model Number when interrogated.
2020-09-28LivaNova USA IncDue to an extra digit being inadvertently added to the serial number of a subset of implantable pulse generators, they become not compatible with therapy programmers with software
2020-02-04LivaNova USA IncFalse positive warning may occur after: 1) VNS Generator interrogated at 0mA normal output current, 2) Generator programmed to non-0mA output current, 3) In-session re-interrogatio
2019-08-22LivaNova USA IncSentiva generators may reset and become disabled within 60 days of enabling. Physicians can re-enable but devices will continue to be susceptible to resets. If a device experiences
2019-08-13LivaNova USA IncFirm identified a subset of its generators that were sterilized one additional sterilization cycle, which does not meet the firm's quality specifications.