Home › THE MAGSTIM COMPANY LIMITED › Neurosign Disposable Twisted Triple Electrodes
Neurosign Disposable Twisted Triple Electrodes
THE MAGSTIM COMPANY LIMITED · Model 1697-00-TE · UDI-DI 05060229530325
10
510(k) clearances
150
Adverse events
2
Recalls
39
Facilities
Description
20mm x 0.3mm needle electrode; 1m cable with TP DIN42802 connector
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2022-08-09 | Technomed Europe | Burns of varying degrees can occur to patients with electrodes left in while in the MRI scanner as a result of inadequate safety information in the IFU. |
| 2007-04-12 | Chalgren Enterprises Inc | Sterility Compromised-- Investigation found that individual packaging seals were separating, which may compromise sterility. |