Recalls

DateFirmReason
2018-03-05Tosoh Bioscience IncAsfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, depending upon the assay.
2016-04-15Siemens Healthcare Diagnostics, Inc.24 minutes of incubation is required prior to processing patient samples and for the pretreatment solution to reach full equilibration when using pretreatment solution when using t
2011-03-30Beckman Coulter Inc.The recall was was initiated because patient results with the IGFBP-3 ELISA kit may be elevated by 70% compared to stated expected ranges in the Instructions for Use (IFU).
2010-08-09Beckman Coulter Inc.The recall was initiated after Beckman Coulter confirmed customer reports that IGFBP-3 measurements using IGFBP-3 IRMA kit (REF DSL-6600) lots identified above may result in IGFBP-
2005-05-02Nichols Institute DiagnosticsUnder recovery of diluted samples by as much as 20%.