Home › DELTEX MEDICAL LIMITED › HD-ICG Electrode Convenience Kits
HD-ICG Electrode Convenience Kits
DELTEX MEDICAL LIMITED · Model 7057-7005 · UDI-DI 05055679404609
DRX
Prescription
10
510(k) clearances
1,587
Adverse events (23 deaths)
20
Recalls
192
Facilities
Description
The CardioQ-EDM+ monitor in HD-ICG mode is intended for use by medically qualified personnel to evaluate and guide optimization of hemodynamic performance in anesthetized, sedated or conscious patients; and is intended for use in the operating room, intensive care unit, emergency room, obstetrics department or other wards or departments where hemodynamic measurement is required. HD-ICG Electrode Convenience Kits contains 2 packs of 3 electrodes and a RFID smart card in a labelled clear plastic bag. The Kits are sold in a labelled shelf box which contains 5 Convenience kits.
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2022-03-15 | Graphic Controls Acquisition Corporation | Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of the standard require |
| 2022-03-15 | Graphic Controls Acquisition Corporation | Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of the standard require |
| 2022-03-15 | Graphic Controls Acquisition Corporation | Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of the standard require |
| 2022-03-15 | Graphic Controls Acquisition Corporation | Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of the standard require |
| 2022-03-15 | Graphic Controls Acquisition Corporation | Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of the standard require |
| 2022-03-15 | Graphic Controls Acquisition Corporation | Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of the standard require |
| 2021-10-06 | 3M Company - Health Care Business | Potential for breakage of the ECG electrode carbon stud from the eyelet. |
| 2020-09-15 | 3M Company - Health Care Business | Corrosion could cause performance failures in the product. |
| 2020-09-15 | 3M Company - Health Care Business | Corrosion could cause performance failures in the product. |
| 2020-09-15 | 3M Company - Health Care Business | Corrosion could cause performance failures in the product. |
Query this across every FDA database — by API.
Linked GUDID · 510(k)/PMA · MAUDE · recalls · establishments. Free tier, no card required.