This product is intended for in vitro diagnostic use in the quality control of IgE on clinical chemistry systems.

Recalls

DateFirmReason
2018-03-05Tosoh Bioscience IncAsfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, depending upon the assay.
2016-09-07Randox Laboratories Ltd.According to the firm, the IgE Calibrator Series IE2492 (lot 387-392IE) value assignment has now been aligned to the 3rd International Standard for human serum IgE 11/234. Turbi
2009-01-05Siemens Medical Solutions DiagnosticsDesign Flaw: The redesign of the Digital Fluidics PCB used in IMMULITE 2000 and IMMULITE 2500 is missing a jumper.