Home › SHANDON DIAGNOSTICS LIMITED › Epredia
Epredia
SHANDON DIAGNOSTICS LIMITED · Model HM525 NX system, 110-120V · UDI-DI 05051663005395
1
510(k) clearances
40
Adverse events
10
Recalls
21
Facilities
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2025-01-30 | LEICA BIOSYSTEMS NUSSLOCH GMBH | Firm became aware of software issues related to most recent firmware update which could result in loss of patient tissue. |
| 2024-09-04 | LEICA BIOSYSTEMS NUSSLOCH GMBH | As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potent |
| 2024-09-04 | LEICA BIOSYSTEMS NUSSLOCH GMBH | As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potent |
| 2024-09-04 | LEICA BIOSYSTEMS NUSSLOCH GMBH | As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potent |
| 2024-09-04 | LEICA BIOSYSTEMS NUSSLOCH GMBH | As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potent |
| 2024-09-04 | LEICA BIOSYSTEMS NUSSLOCH GMBH | As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potent |
| 2024-09-04 | LEICA BIOSYSTEMS NUSSLOCH GMBH | As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potent |
| 2019-05-03 | Leica Microsystems, Inc. | The current labelling does not include any specific warning about usage of flammable freezing sprays. Such flammable freezing sprays can potentially ignite and can cause serious i |
| 2010-09-23 | Thermo Fisher Scientific | Through internal quality review, firm determined the need to conduct a more comprehensive system validation to support the claim of complete bacterial, fungal and viral disinfectio |
| 2010-09-23 | Thermo Fisher Scientific | Through internal quality review, firm determined the need to conduct a more comprehensive system validation to support the claim of complete bacterial, fungal and viral disinfectio |
Query this across every FDA database — by API.
Linked GUDID · 510(k)/PMA · MAUDE · recalls · establishments. Free tier, no card required.
Get your free API keyRead the docsData compiled from public U.S. FDA sources (GUDID, 510(k)/PMA, MAUDE, recalls, establishment registration). Adverse-event reports are unverified and do not establish causation. Informational only — not medical, clinical, or regulatory advice.
