Home › NIDEK CO.,LTD. › NA
NA
NIDEK CO.,LTD. · Model EC-5000 · UDI-DI 04987669100011
LZS
Prescription
0
510(k) clearances
13,741
Adverse events (12 deaths)
29
Recalls
10
Facilities
Description
NIDEK EXCIMER LASER CORNEAL SURGERY SYSTEM, model EC-5000CXIII, is an ophthalmic laser system used for correction of corneal refraction and ablation of the corneal surface. The system is composed of a laser generator which produces an excimer laser radiation of wavelength 193nm, a beam delivery unit, an optical system for observation, a gas system, and a computer for system control. The system is designed to be user friendly for easy and safe operation with NIDEK’s advanced optoelectronic technology.
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2021-10-21 | AMO Manufacturing USA, LLC | Aberrometer and precision laser systems are affected by a remote code execution vulnerability existing when the Microsoft Windows Print Spooler service is enabled, and improperly p |
| 2017-02-01 | Nidek Inc | During treatment planning, the procedure was programmed with an unintended (wrong) correction. |
| 2015-03-24 | Nidek Inc | Multi-stage treatment option for the EC-5000 Operator's Manual was not approved by the FDA. |
| 2013-09-16 | Abbott Medical Optics, Inc. | This corrective action will address an identified failure of the lamp and illumination system within the laser unit caused by an unexpected failure of a thyristor component on the |
| 2013-09-16 | Abbott Medical Optics, Inc. | This corrective action will address an identified failure of the lamp and illumination system within the laser unit caused by an unexpected failure of a thyristor component on the |
| 2013-09-16 | Abbott Medical Optics, Inc. | This corrective action will address an identified failure of the lamp and illumination system within the laser unit caused by an unexpected failure of a thyristor component on the |
| 2013-09-16 | Abbott Medical Optics, Inc. | This corrective action will address an identified failure of the lamp and illumination system within the laser unit caused by an unexpected failure of a thyristor component on the |
| 2013-05-28 | Abbott Medical Optics, Inc. | Under specific conditions, there is an unlikely potential to result in inadvertent laser firing of the laser. |
| 2013-05-28 | Abbott Medical Optics, Inc. | Under specific conditions, there is an unlikely potential to result in inadvertent laser firing of the laser. |
| 2013-05-28 | Abbott Medical Optics, Inc. | Under specific conditions, there is an unlikely potential to result in inadvertent laser firing of the laser. |
Query this across every FDA database — by API.
Linked GUDID · 510(k)/PMA · MAUDE · recalls · establishments. Free tier, no card required.
Get your free API keyRead the docsData compiled from public U.S. FDA sources (GUDID, 510(k)/PMA, MAUDE, recalls, establishment registration). Adverse-event reports are unverified and do not establish causation. Informational only — not medical, clinical, or regulatory advice.