FDA Device Intelligence
Home › SANTEC CORPORATION › OPTICAL BIOMETER

OPTICAL BIOMETER

SANTEC CORPORATION · Model ARGOS · UDI-DI 04573344510017

MXK Prescription
1
510(k) clearances
12
Adverse events
7
Recalls
18
Facilities

Description

ARGOS is a non-invasive, non-contact biometer based on swept-source optical coherence tomography (SS-OCT). The device is intended to acquire ocular measurements as well as perform calculations to determine the appropriate intraocular lens (IOL) power and type for implantation during intraocular lens placement. ARGOS measures the following 9 parameters: Axial Length, Corneal Thickness, Anterior Chamber Depth, Lens Thickness, K-values (Radii of flattest and steepest meridians), Astigmatism, Whiteto- white (corneal diameter) and Pupil Size. It is intended for use by ophthalmologists, physicians, and other eye-care professionals and may only be used under the supervision of a physician.

Recalls

DateFirmReason
2023-08-08Oculus Optikgeraete GMBHTheir is a potential that optical devices with insufficient anti-reflective coating may lead to incorrect axial length measurements.
2023-08-08Oculus Optikgeraete GMBHTheir is a potential that optical devices with insufficient anti-reflective coating may lead to incorrect axial length measurements.
2023-08-08Oculus Optikgeraete GMBHTheir is a potential that optical devices with insufficient anti-reflective coating may lead to incorrect axial length measurements.
2022-07-08Oculus Optikgeraete GMBHDue to a software issue, IOL calculator printout often does not accurately reflect the alignment axis and incision position when planning toric IOLs
2022-07-08Oculus Optikgeraete GMBHDue to a software issue, IOL calculator printout often does not accurately reflect the alignment axis and incision position when planning toric IOLs
2022-07-08Oculus Optikgeraete GMBHDue to a software issue, IOL calculator printout often does not accurately reflect the alignment axis and incision position when planning toric IOLs
2017-11-21Oculus Optikgeraete GMBHThe device software versions have an anomaly which may produce an erroneous marking for the quality specification value

Query this across every FDA database — by API.

Linked GUDID · 510(k)/PMA · MAUDE · recalls · establishments. Free tier, no card required.

Get your free API keyRead the docs
Data compiled from public U.S. FDA sources (GUDID, 510(k)/PMA, MAUDE, recalls, establishment registration). Adverse-event reports are unverified and do not establish causation. Informational only — not medical, clinical, or regulatory advice.