Facilities
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2024-10-09 | Shimadzu Corp. Analytical & Measuring Instruments Div. | There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping ont |
| 2024-10-09 | Shimadzu Corp. Analytical & Measuring Instruments Div. | There have been cases of failure of the main board due to condensation water drained from the Peltier dehumidifier, leaked out from the tray, resulting in a short circuit. In the s |
Query this across every FDA database — by API.
Linked GUDID · 510(k)/PMA · MAUDE · recalls · establishments. Free tier, no card required.
Get your free API keyRead the docsData compiled from public U.S. FDA sources (GUDID, 510(k)/PMA, MAUDE, recalls, establishment registration). Adverse-event reports are unverified and do not establish causation. Informational only — not medical, clinical, or regulatory advice.