FDA Device Intelligence
Home › Tem Innovations GmbH › ROTEM®

ROTEM®

Tem Innovations GmbH · Model hep-tem® · UDI-DI 04260160470082

JPAGGN
1
510(k) clearances
7,801
Adverse events (39 deaths)
43
Recalls
41
Facilities

Description

The ROTEM® delta is designed for in-vitro diagnostic use by professionals in a laboratory environ-ment. The ROTEM® delta is intended to be used to provide a qualitative and quantitative indication of the coagulation state of a blood sample. For this purpose the ROTEM® delta records the clot firmness changes in a sample of citrated whole blood as the sample clots, retracts and lyses in real time. The analyzer output consists of a qualitative graphical representation (mirrored coagulation curve - clot firmness over time) and several defined numerical parameters describing the curve quantitatively. The hep-tem® assay is a semi-quantitative in vitro diagnostic assay used to inactivate heparin in patients receiving unfractionated heparin on the ROTEM® delta. To monitor the coagulation process citrated whole blood specimens are used. Clotting characteristics are described by the functional parameters Clotting Time (CT), Speed of Clot Formation (CFT and alpha angle), Clot Firmness (A20/MCF) and Clot Lysis (LOT, ML, LI(x)).

Recalls

DateFirmReason
2025-07-21Diagnostica Stago, Inc.Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.
2023-03-22Diagnostica Stago, Inc.An internal investigation identified a bug in this firmware version, resulting in the following:Intermittent shortened coagulation times and an increased frequency of occurrence of
2023-03-22Diagnostica Stago, Inc.An internal investigation identified a bug in this firmware version, resulting in the following:Intermittent shortened coagulation times and an increased frequency of occurrence of
2023-03-22Diagnostica Stago, Inc.An internal investigation identified a bug in this firmware version, resulting in the following:Intermittent shortened coagulation times and an increased frequency of occurrence of
2022-04-26Haemonetics CorporationDue to a shift in the citrated blood reference range for a normal population, the ranges stated in the IFU are changing. If the clinician relies solely on affected results in compa
2022-04-19Haemonetics CorporationWhen the TEG 5000 Analyzer including TEG Analytical Software is used with PlateletMapping (ADP or AA) Assays and TEG Mangement Software, incorrect values of the PlateletMapping ADP
2021-03-02Siemens Healthcare Diagnostics, Inc.Siemens Healthcare Diagnostics Products GmbH has confirmed that the BCS XP System may be affected by a potential Emicizumab carryover on patient samples.
2020-11-09Helena Laboratories, Corp.Due to complaints received associated with cracked/split test tube caps.
2020-02-19Haemonetics CorporationSoftware defect in TEG Manager impacts the displayed alert for out of range test results. Due to this defect, reference range values received from TEG 5000 and displayed on the TEG
2019-07-23Instrumentation Laboratory Co.Prolonged clotting times

Query this across every FDA database — by API.

Linked GUDID · 510(k)/PMA · MAUDE · recalls · establishments. Free tier, no card required.

Get your free API keyRead the docs
Data compiled from public U.S. FDA sources (GUDID, 510(k)/PMA, MAUDE, recalls, establishment registration). Adverse-event reports are unverified and do not establish causation. Informational only — not medical, clinical, or regulatory advice.