AquaBplus
VIVONIC GmbH · Model AquaBplus HF · UDI-DI 04251285200296
Description
Additional flow heater unit for AquaBplus Water Treatment Systems. AquaBplus HF Unit for: AquaBplus & AquaBplus HF 500 (Primary DI No.: 04251285200128) AquaBplus & AquaBplus HF 1000 (Primary DI No.: 04251285200135) AquaBplus & AquaBplus HF 1500 (Primary DI No.: 04251285200142) AquaBplus & AquaBplus HF 2000 (Primary DI No.: 04251285200159) AquaBplus & AquaBplus HF 2500 (Primary DI No.: 04251285200166) AquaBplus & AquaBplus HF 3000 (Primary DI No.: 04251285200173) AquaBplus & AquaBplus B2 & AquaBplus HF 500 (Primary DI No.: 04251285200180) AquaBplus & AquaBplus B2 & AquaBplus HF 1000 (Primary DI No.: 04251285200197) AquaBplus & AquaBplus B2 & AquaBplus HF 1500 (Primary DI No.: 04251285200203) AquaBplus & AquaBplus B2 & AquaBplus HF 2000 (Primary DI No.: 04251285200210) AquaBplus & AquaBplus B2 & AquaBplus HF 2500 (Primary DI No.: 04251285200227)
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2024-05-16 | Fresenius Medical Care Holdings, Inc. | Concentrate Flow was exceeding the maximum 19.9 bar set point. The standard default set point range is between 4 and 18.Error message (F020104, "concentrate pressure alarm limit ex |
| 2023-01-24 | Outset Medical, Inc. | An observed trend of high conductivity dialysate alarms in a hemodialysis system as a result of introducing a software tool into the manufacturing process. |
| 2022-06-21 | Fresenius Medical Care Holdings, Inc. | Internal ethemet cables are missing one or both ferrite cores therefore the electromagnetic compatibility of the device is not ensured |
| 2022-06-21 | Fresenius Medical Care Holdings, Inc. | Internal ethemet cables are missing one or both ferrite cores therefore the electromagnetic compatibility of the device is not ensured |
| 2022-06-16 | Fresenius Medical Care Holdings, Inc. | Internal ethernet cable may be misaligned, the cable jacket (insulation) could become compromised due to high temperature and result in the internal ethernet cable conducting elect |
| 2022-06-16 | Fresenius Medical Care Holdings, Inc. | Internal ethernet cable may be misaligned, the cable jacket (insulation) could become compromised due to high temperature and result in the internal ethernet cable conducting elect |
| 2021-08-03 | Fresenius Medical Care Holdings, Inc. | AquaBplus HF module on the AquaBplus 2000 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combi |
| 2021-03-18 | Fresenius Medical Care Holdings, Inc. | AquaBplus HF module on the AquaBplus 2500 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combi |
| 2020-09-01 | Fresenius Medical Care Holdings, Inc. | Incorrectly crimped connectors were installed on the main protective earth conductor on these machines. If the machine is moved or maintenance is performed, these crimp connections |
| 2014-04-24 | Mar Cor Purification | Mar Cor Purification manufactured Central Water Plant (CWP) reverse osmosis system, models 102, 104 and 106. This correction has been initiated due to a potential situation with a |