FDA Device Intelligence
Home › Aesku.Diagnostics GmbH & Co. KG › AESKULISA® ANA-HEp-2

AESKULISA® ANA-HEp-2

Aesku.Diagnostics GmbH & Co. KG · Model 7115US · UDI-DI 04250289503020

DHN
1
510(k) clearances
1
Adverse events
7
Recalls
19
Facilities

Description

AESKULISA ANA-HEp-2 is a solid phase enzyme immunoassay for the combined qualitative detection of IgG antibodies against HEp2 cells in human serum. Each well is coated with lysed HEp2 cells. The test collectively detects, in one well, total ANAs against double stranded DNA (dsDNA), histones, SS-A (Ro), SS-B (La), Sm, snRNP/Sm, Scl-70, PM-Scl, Jo-1 and centromeric antigens along with sera positive for HEp2 immunofluorescence test (IFT). The assay is a tool in the diagnosis of systemic rheumatic diseases like systemic lupus erythematosus (SLE), mixed connective tissue diseases (MCTD), scleroderma, Sjögren`s syndrome, polymyositis and dermatomyositis.

Recalls

DateFirmReason
2011-09-26Inova Diagnostics IncorporatedThe NOVA Lite ANA KSL kit label has the wrong fill volume for the IFA System Negative Control (P/N 508186) with 1mL. The component in the kit is correctly labeled and filled with
2010-06-30Bio-Rad LaboratoriesKALLESTAD Mouse Stomach/Kidney (MSK) test kits contained Crithidia slides instead of the MSK slides.
2008-12-04Immuno Concepts IncSamples can occasionally run out of the wells and cross contaminate adjacent wells.
2008-07-29Bio-Rad LaboratoriesThe Homogeneous Positive control was released with a titer of 1:256 rather than the titer value of 1:128 reported on the QC Report Label included with each test kit.
2004-09-24Diasorin Inc.A problem with software for the PolyTiter Immunofluorescent Titration System in that no flags or warnings appear when PolyTiter Calibrator Values are re-entered or changed (includi
2004-06-30Diasorin Inc.The program for a specific immunofluorescent titration system (Polymedco PolyTiter immunofluorescent Titration System) was found faulty in that the calibrator 2 value was incorrect
2003-07-21Perkinelmer Life Sciences, Inc.Product's outer carton label was labeled with an extended expiration date

Query this across every FDA database — by API.

Linked GUDID · 510(k)/PMA · MAUDE · recalls · establishments. Free tier, no card required.

Get your free API keyRead the docs
Data compiled from public U.S. FDA sources (GUDID, 510(k)/PMA, MAUDE, recalls, establishment registration). Adverse-event reports are unverified and do not establish causation. Informational only — not medical, clinical, or regulatory advice.