FDA Device Intelligence
Home › Brainlab SE › Automatic Registration

Automatic Registration

Brainlab SE · Model 2.6 · UDI-DI 04056481145101

OLOHAW Prescription
3
510(k) clearances
20,634
Adverse events (13 deaths)
114
Recalls
383
Facilities

Description

The Automatic Registration - is a software module, that when used with a compatible computer workstation and compatible registration accessories provides a “registration” for subsequent navigation applications.

Recalls

DateFirmReason
2026-08-28Intellijoint Surgical, Inc.A navigation system used to aid in selection and positioning of orthopedic implant system components has a workstation that may contain a 15-inch MacBook Pro battery, subject to an
2026-07-23Medos International SarlSoftware issue within robotic controller which may cause unintended movement of the robotic arm during bone preparation or instrument positioning.
2026-05-08Think Surgical, Inc.Due to manufacturing/design issue where the knob on the tracking array may loosen causing potential misalignment of the array.
2026-04-23Kico Knee Innovation CompanyThe Impactor V-Block Assembly and Trackers A, B, C and E may exhibit signs of early wear and/or corrosion on the exposed magnet face.
2026-04-23Kico Knee Innovation CompanyThe V-Block Assembly and Trackers A, B, C and E may exhibit signs of early wear and/or corrosion on the exposed magnet face.
2026-02-10Orthalign, IncDrill plates manufactured out of specification can result in a bone cut in excess 00800"(2.8448mm)
2026-02-10Orthalign, IncDrill plates manufactured out of specification can result in a bone cut in excess 00800"(2.8448mm)
2026-02-02Howmedica Osteonics Corp.A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw
2026-01-02Alphatec Spine, Inc.Due a design issue where the navigated array connection geometry is incorrect.
2026-01-02Alphatec Spine, Inc.Due a design issue where the navigated array connection geometry is incorrect.