therascreen® FGFR RGQ RT-PCR Kit
QIAGEN GmbH · Model 2 · UDI-DI 04053228043560
Description
The therascreen FGFR RGQ RT-PCR Kit is a real-time, reverse transcription PCR test for the qualitative detection of two point mutations in exon 7 [p.R248C (c.742C>T), p.S249C (c.746C>G)], two point mutations in exon 10 [p.G370C (c.1108G>T) and p.Y373C (c.1118A>G)] and two fusions (FGFR3-TACC3v1 and FGFR3-TACC3v3) in the fibroblast growth factor receptor 3 (FGFR3) gene in RNA samples derived from formalin-fixed paraffin-embedded (FFPE) urothelial tumor tissue. The test is indicated for use as an aid in identifying patients with cases of urothelial cancer (UC) which harbor these alterations and are therefore eligible for treatment with BALVERSA™ (erdafitinib). Specimens are processed using the RNeasy DSP FFPE Kit for manual sample preparation followed by reverse transcription and then automated amplification and detection on the Rotor-Gene Q MDx (US) instrument.
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2023-08-22 | Qiagen GmbH | Handbook Revision (Rev13) a dilution step is indicated as an Action when having the Flag SAMPLE_INT_CTRL_FAIL. A dilution of CRC tissue could lead to a false negative result. To r |
| 2022-07-20 | Qiagen Sciences LLC | Increased occurrence rate of false positive results, lead to a false positive sample result that could subsequently be incorrectly reported from the laboratory. A false positive re |
| 2022-07-20 | Qiagen Sciences LLC | Increased occurrence rate of false positive results, lead to a false positive sample result that could subsequently be incorrectly reported from the laboratory. A false positive re |
| 2022-07-20 | Qiagen Sciences LLC | Increased occurrence rate of false positive results, lead to a false positive sample result that could subsequently be incorrectly reported from the laboratory. A false positive re |
| 2022-02-25 | Qiagen Sciences LLC | False positive or false negative G12C (12CYS) mutation result for the KRAS Mutation Status resulting from Flags/Warnings may invalidate the result for one or more individual mutati |
| 2022-02-25 | Qiagen Sciences LLC | False positive or false negative G12C (12CYS) mutation result for the KRAS Mutation Status resulting from Flags/Warnings may invalidate the result for one or more individual mutati |
| 2021-09-03 | Roche Molecular Systems, Inc. | False Mutation Detected results for the exon 20 insertion (Ex20Ins) mutation when using the cobas EGFR Mutation Test v2, may lead to inappropriate treatment |
| 2020-12-11 | Qiagen Sciences LLC | Kit may generate false Q546R mutation positive results caused by nonspecific molecular interactions within the Q546R higher than previously observed and described within the Instru |
| 2016-04-05 | Roche Molecular Systems, Inc. | During internal studies using contrived plasma specimens, several mutations (L8568R, Exon 19 deletion, T790M) inconsistently generated "No Mutation Detected" (ie False Negative ) r |
| 2016-03-15 | Roche Molecular Systems, Inc. | An error was found within the Hungarian translations of the cobas EGFR Mutation Test v2 Instructions for Use (M/N 07340761001-01HU, Doc Rev. 1.0, Dated 08/2015) and the cobas cfDNA |