therascreen® PIK3CA RGQ PCR Kit
QIAGEN GmbH · Model V1 · UDI-DI 04053228034223
Description
The therascreen PIK3CA RGQ PCR Kit is a real-time qualitative PCR test for the detection of 11 mutations in the phosphatidylinositol 3-kinase catalytic subunit alpha (PIK3CA) gene (Exon 7: C420R; Exon 9: E542K, E545A, E545D [1635G>T only], E545G, E545K, Q546E, Q546R; and Exon 20: H1047L, H1047R, H1047Y) using genomic DNA (gDNA) extracted from formalin-fixed, paraffin-embedded (FFPE) breast tumor tissue or circulating tumor DNA (ctDNA) from plasma derived from K2EDTA anticoagulated peripheral whole blood taken from patients with breast cancer. The test is intended to aid clinicians in identifying breast cancer patients who may be eligible for treatment with PIQRAY® (alpelisib) based on a PIK3CA Mutation Detected result. Patients whose FFPE tissue or plasma specimen produces a positive therascreen PIK3CA RGQ PCR Kit test result for the presence of one or more PIK3CA mutations are eligible for treatment with PIQRAY (alpelisib). Patients whose plasma specimen produces a negative result using this test should be reflexed to testing with FFPE tumor tissue for the presence of PIK3CA mutations. FFPE tumor specimens are processed using the QIAamp DSP DNA FFPE Tissue Kit for manual sample preparation. K2EDTA anticoagulated whole peripheral venous blood plasma specimens are processed using the QIAamp DSP Circulating Nucleic Acid Kit for manual sample preparation. For both specimen types, the Rotor-Gene Q (RGQ) MDx (US) instrument is used for automated amplification and detection. The Kit is to be used by trained personnel in a professional laboratory environment.
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2023-08-22 | Qiagen GmbH | Handbook Revision (Rev13) a dilution step is indicated as an Action when having the Flag SAMPLE_INT_CTRL_FAIL. A dilution of CRC tissue could lead to a false negative result. To r |
| 2022-07-20 | Qiagen Sciences LLC | Increased occurrence rate of false positive results, lead to a false positive sample result that could subsequently be incorrectly reported from the laboratory. A false positive re |
| 2022-07-20 | Qiagen Sciences LLC | Increased occurrence rate of false positive results, lead to a false positive sample result that could subsequently be incorrectly reported from the laboratory. A false positive re |
| 2022-07-20 | Qiagen Sciences LLC | Increased occurrence rate of false positive results, lead to a false positive sample result that could subsequently be incorrectly reported from the laboratory. A false positive re |
| 2022-02-25 | Qiagen Sciences LLC | False positive or false negative G12C (12CYS) mutation result for the KRAS Mutation Status resulting from Flags/Warnings may invalidate the result for one or more individual mutati |
| 2022-02-25 | Qiagen Sciences LLC | False positive or false negative G12C (12CYS) mutation result for the KRAS Mutation Status resulting from Flags/Warnings may invalidate the result for one or more individual mutati |
| 2021-09-03 | Roche Molecular Systems, Inc. | False Mutation Detected results for the exon 20 insertion (Ex20Ins) mutation when using the cobas EGFR Mutation Test v2, may lead to inappropriate treatment |
| 2020-12-11 | Qiagen Sciences LLC | Kit may generate false Q546R mutation positive results caused by nonspecific molecular interactions within the Q546R higher than previously observed and described within the Instru |
| 2016-04-05 | Roche Molecular Systems, Inc. | During internal studies using contrived plasma specimens, several mutations (L8568R, Exon 19 deletion, T790M) inconsistently generated "No Mutation Detected" (ie False Negative ) r |
| 2016-03-15 | Roche Molecular Systems, Inc. | An error was found within the Hungarian translations of the cobas EGFR Mutation Test v2 Instructions for Use (M/N 07340761001-01HU, Doc Rev. 1.0, Dated 08/2015) and the cobas cfDNA |