FDA Device Intelligence
Home › Leica Biosystems Nussloch GmbH › Bath insert Medite/Hacker 30

Bath insert Medite/Hacker 30

Leica Biosystems Nussloch GmbH · Model 000000014047837263 · UDI-DI 04049188052681

KIM
1
510(k) clearances
10
Adverse events
1
Recalls
13
Facilities

Description

Leica Biosystems Accessory

Recalls

DateFirmReason
2018-09-04Leica Microsystems, Inc.Product was assembled with an isolating fiberglass hose within the oven in the wrong position. This results in improper shielding from potential electrical shock.

Query this across every FDA database — by API.