DIPLOMAT
SIGNUS Medizintechnik GmbH · Model Pedicle screw w/o tulip, fenestrated, polyaxial · UDI-DI 04047844019443
Description
The modular DIPLOMAT® pedicle screw system is a rod-screw system that firmly connects two or more screws via the screw head and a rod using a force-fit or form-fit connection. Additional fixation elements, such as cross bars, hooks, connectors and washers, can also be used. The DIPLOMAT® pedicle screw system is applied temporarily to provide internal posterior stabilization until bone fusion in the lumbar or thoracic spine has taken place. The surgeon makes the final decision as to the indwelling time of the implant in the patient‘s body and the time when the implant is to be explanted. The pedicle screw system is implanted from a posterior approach. The implant is screwed into the vertebral body via the pedicle and it can be applied in a single segment or multiple segments. Implantation is facilitated by use of the specially developed accessories for inserting and positioning the implants. Only these accessories ensure safe use. Our product information provides further system-related information on the surgical method. If there are any preoperative uncertainties relating to the implant, information must be obtained from SIGNUS. Instruments specially developed by SIGNUS to ensure safe application are available for use with the implant system.
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2017-04-12 | XTANT MEDICAL INC | The material type on the label may incorrectly state Ti6Al-4V ELI. The rods are composed of Cobalt Chromium. |
| 2016-11-22 | Amendia, Inc | The rods are packaged in packages with unacceptable seals. In some cases, the integrity of the seal is compromised by channels or voids. |
| 2015-04-07 | X Spine Systems Inc | The Continuous Radius Rod should measure 50mm in length but actually measures 45mm in length. |
| 2015-03-11 | X Spine Systems Inc | One lot of instruments have etched on the Yoke of the Pedicle Screw a laser marking size of 7.5 x 55mm instead of 5.5 x 55mm as described on the product label. |
| 2013-01-11 | Synthes USA HQ, Inc. | Lot number 7542279, of the 45 mm Click 'X Pedicle Screw (Part number 487.072), was incorrectly etched as "Click 'X Pedicle Screw 05.2 L40 TAN violet" although the actual length of |
| 2012-09-25 | Alphatec Spine, Inc. | The firm initiated the recall because the stainless steel guide-wire of the Illico MIS Fixation System did not completely fit/pass through the tip of certain Illico Ti Cannulated P |
| 2009-07-01 | Stryker Spine | The Trio 3.5MM Hex socket has been reported to break under high stresses. |
| 2009-04-30 | Stryker Spine | The Mantis Rod Inserter Shaft was found to have deformation of the stainless stell split sphere when subjected to high forces, prohibiting the user from being able to grip the rod |
| 2009-04-30 | Stryker Spine | The Mantis Rod Inserter Shaft was found to have deformation of the stainless stell split sphere when subjected to high forces, prohibiting the user from being able to grip the rod |
| 2009-04-02 | EBI, L.P. | Some devices could have been made with raw material lots that may have inclusions, and could result in decreased strength of the device. |