Home › Aesculap AG › AESCULAP
AESCULAP
Aesculap AG · Model FT502T · UDI-DI 04046963386320
HCI
Prescription
10
510(k) clearances
54
Adverse events
5
Recalls
10
Facilities
Description
YASARGIL Clip Applier Forceps, bayonet, 9 3/4", (245 mm), working length: 4 3/8", (110 mm), bayonet-shaped, with ratchet, standard,stainless steel, non-sterile, reusable
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2023-10-06 | Peter Lazic Gmbh | Customer complaint received that reported the applier failed during surgery. The applier did not release the clip. The applier remained attached to the clip. |
| 2023-10-06 | Peter Lazic Gmbh | Customer complaint received that reported the applier failed during surgery. The applier did not release the clip. The applier remained attached to the clip. |
| 2023-10-06 | Peter Lazic Gmbh | Customer complaint received that reported the applier failed during surgery. The applier did not release the clip. The applier remained attached to the clip. |
| 2007-11-01 | Aesculap Inc | Breakage: the inner rod of the applier may corrode and break |
| 2007-11-01 | Aesculap Inc | Breakage: the inner rod of the applier may corrode and break |