Home › DIABETIC SUPPLY OF SUNCOAST, INC. › Advocate Lancets Safety 28 Gauge
Advocate Lancets Safety 28 Gauge
DIABETIC SUPPLY OF SUNCOAST, INC. · Model 637 · UDI-DI 00894046001912
10
510(k) clearances
14,134
Adverse events (2 deaths)
17
Recalls
164
Facilities
Description
Advocate Lancets Safety 28 Gauge 1.8 mm box of 200, sterile, for capillary blood sample, needle automatically retracts, ready to use, no loading needed, single use only
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2021-03-16 | Cardinal Health Inc. | Due to shipping delays the devices were exposed to temperature excursions of 31.9F, intermittently over a 2 hour and 15 minute period of time which may cause inaccurate results. |
| 2021-03-16 | Cardinal Health Inc. | Due to shipping delays the devices were exposed to temperature excursions of 31.9F, intermittently over a 2 hour and 15 minute period of time which may cause inaccurate results. |
| 2021-03-16 | Cardinal Health Inc. | Due to shipping delays the devices were exposed to temperature excursions of 31.9F, intermittently over a 2 hour and 15 minute period of time which may cause inaccurate results. |
| 2021-03-16 | Cardinal Health Inc. | Due to shipping delays the devices were exposed to temperature excursions of 31.9F, intermittently over a 2 hour and 15 minute period of time which may cause inaccurate results. |
| 2013-08-19 | Roche Diagnostics Operations, Inc. | Internal Roche inspections have revealed that after the lancet is used for a finger stick, the needle may not fully retract back within the lancet drum or back into the cap. This a |
| 2013-07-10 | Becton Dickinson & Company | The Lancet will not retract after activation thus creating a potential for exposure to a used sharp. |
| 2013-07-03 | KooJoo Trading Company | Lancets were not sterilized within the dose range established by sterilization validation. |
| 2009-05-13 | Roche Diagnostics Corp. | The lancet cap may be missing, which could result in an unintended lancet stick to the user. |
| 2008-08-27 | Sarstedt Inc | The product from this lot may not retract the lancet into the safe position inside the lancet body after the trigger is actuated. |
| 2008-07-18 | Facet Technologies | The Alternate Site Cap was not included in the package as indicated on the labeling. |