Life Warmer Quantum Battery
NORTH AMERICAN RESCUE, LLC · Model 63-5902 · UDI-DI 00868058000423
Description
The LifeWarmer Quantum Battery is an extremely powerful and intelligent battery in a small, lightweight package- weighing only 16 oz. The Quantum Battery powers the whole warming system with state-of-the-art Lithium Polymer cells harnessing 44 Volts of electrical potential. The battery, fully charged, is able to warm and maintain up to 2 units ( 900mL +/- 100mL) of cold (4C) blood products or IV solutions to a pre-set temperature of 100.4°F (38+/- 2C) at a flow rate of 100 mL/min (depending on the ambient temperature). Or, approximately 1700 mL of 20C IV solution to a set point of 38C +/- 2C at a flow rate of 200 mL/min. The Quantum Battery is built for use in the most austere and tactical environments. It is IP67 rated for water resistance and is the only portable fluid warming battery that has a disable feature on the audible alarm for tactical considerations. The Quantum Life Warmer Battery is rechargeable with a service life of 500 charge/discharge cycles. It is important to keep the Quantum Battery charged and ready for use. Always re-charge after deployment, even if the battery was not fully discharged. The battery can be kept plugged into the Quantum Charger, as the intelligent Battery senses charge demand and will charge as needed but never overcharge. The Battery has a “Status” button to check the state of charge at any moment, with 3 colored LEDs to inform the user about the condition of the battery.
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2026-06-24 | ICU Medical Inc. | ICU Medical became aware of an increased likelihood of leakage in specific Lots of L-70 HOTLINE Fluid Warming Sets due to potentially inadequate solvent application at the tubing t |
| 2025-04-22 | 3M Company | Notice was issued to clarify flow rates related to the inlet fluid temperature. |
| 2025-03-31 | Belmont Instrument LLC | Potential crack in the female connector located on the Heat Exchanger in the set, which will lead to a fluid leak through the crack during priming of the rapid infuser. |
| 2025-03-31 | Belmont Instrument LLC | Potential crack in the female connector located on the Heat Exchanger/patient line set, which will lead to a fluid leak through the crack during priming of the Rapid Infuser. |
| 2025-03-31 | Belmont Instrument LLC | Potential crack in the female connector located on the Heat Exchanger in the set, which will lead to a fluid leak through the crack during priming of the rapid infuser. |
| 2025-03-31 | Belmont Instrument LLC | Potential crack in the female connector located on the Heat Exchanger in the set, which will lead to a fluid leak through the crack during priming of the rapid infuser. |
| 2021-08-06 | Smiths Medical ASD Inc. | There is potential for harmful levels of aluminum leaching when using lower flow rates with certain solutions and blood products and longer duration of use. |
| 2021-08-06 | Smiths Medical ASD Inc. | There is potential for harmful levels of aluminum leaching when using lower flow rates with certain solutions and blood products and longer duration of use. |
| 2021-06-04 | Eight Medical International BV | The patient may be exposed to potential aluminum ion release during warming therapy under certain use conditions. |
| 2021-04-30 | RanD S.r.l. | The sterilization contractor informed the company of a quality issue possibly affecting the sterilization process of a certain number of products. It was determined that some lots |