FDA Device Intelligence
Home › Lucas Lifecare, LLC › Lucas Lifecare

Lucas Lifecare

Lucas Lifecare, LLC · Model LifeUltra · UDI-DI 00860014827310

FLE
1
510(k) clearances
1,513
Adverse events (1 deaths)
18
Recalls
61
Facilities

Description

Life Ultra 25L Tabletop Mono Chamber Steam Sterilizer is a fully automatic tabletop steam sterilizer that uses saturated steam under pressure to sterilize heat- and moisture-resistant reusable loads (solid, hollow, porous, or delicate) in wrapped or unwrapped configurations for medical, dental, surgical, healthcare, and laboratory settings. The unit incorporates a vacuum system for dynamic air removal to support steam penetration, temperature uniformity, and post-sterilization drying (drying with door closed). The sterilizer is microcontroller-controlled with installed software and pre-set sterilization programs, providing validated Class B and Class S cycles along with test cycles including Vacuum Test and Bowie-Dick Test.

Recalls

DateFirmReason
2026-07-08Fort Defiance Industries, LLCThe Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable f
2023-06-22Scican Ltd.The STATCLAVE G4 is a dynamic air removal (prevacuum) tabletop steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized ste
2023-06-22Scican Ltd.The STATCLAVE G4 is a dynamic air removal (prevacuum) tabletop steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized ste
2021-08-19Fort Defiance Industries, LLCThe rear heater plate and cover assembly that contains the electrical wiring for the heaters does not have sufficient bonding to ground.
2019-10-22Steris CorporationPotential for the incorrect pressure safety valve being included in the Preventive Maintenance (PM) kit for the 3000 Small Sterilizers.
2019-10-22Steris CorporationThe locking pin of the front wheel asse mbly of the Evolution Transfer Carriages may not remain full y engaged with the transfer carriage and result in the transfer carriage and lo
2019-04-18Skytron, Div. The KMW Group, IncWeld attaching the air heater assembly to the main sterilizer unit has shown signs of fatigue/failure
2018-12-26Steris CorporationThe ATLAS loading car may not remain fully engaged with the latch on the transfer carriage when the loading car is empty, potentially resulting in the loading car unexpectedly slid
2018-05-24Fort Defiance Industries, LLCDuring long-term storage (i.e. military depot), the interaction of stagnant water and dissimilar metals in the plumbing of the P2131 device produces deposits in the water system th
2015-09-03Steris CorporationIn AMSCO Small Steam Sterilizers equipped with the CS-iQ feature, the sterilizer software inadvertently prevents the capability to operate the foot pedal, which allows the Operator

Query this across every FDA database — by API.